67
5.
Open the door.
CAUTION! The rack and chamber may be hot if the unit is in standby mode.
6.
Insert the temperature probe through the temperature port (TWINLOK001) on the calibration fitting
making sure the tip of the probe is well inside the chamber and not touching the rack.
7.
Switch on both the temperature and pressure reference meters. Ensure the pressure meter is set to
kPa
and the temperature meter is in
C°
.
8.
Close the chamber door.
9.
Connect one end of the pressure meter tube to your reference meter.
10.
Connect the other end of the pressure meter tube to the pressure port on the calibration fitting.
STAGE 2. Starting calibration
11.
From the STATCLAVE G4 home screen, access the calibration function by following these steps:
Option 1 (as provided)
Option 2 (assembly required)
Teflon washer
Knurled nut
pressure port
Temperature port
Quick connect
pressure port
Teflon washer
Temperature port
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...