69
To complete the
POA
calibration, compare the pressure reading on the display (
P
) with the pressure reading
on the pressure reference meter. (See the following Step for a full explanation of the calibration screen.)
Adjust the pressure reading using the up or down arrows on the
POA
offset field.
NOTE: The difference in two readings should be within 0.5Kpa. If the two readings cannot be matched
precisely, set pressure reading on unit to one count below the reading on the pressure reference
meter.
13.
Explanation of the calibration screen:
14.
After completing the POA calibration, use the arrows below the
POS
and
TO
values to set them to
00
,
if the displayed values are not already 00.
NOTE: Offset adjustment arrows DON’T have to be highlighted to change values.
STAGE 4. Adjusting the pressure at vacuum
15.
The cycle then continues to the vacuum phase. The
POV
field on the display will be highlighted.
16.
Wait for the vacuum to reach
60 Kpa or less
. Once the vacuum is in this range, compare the reading
on unit’s display (
P
)
with the reading on pressure reference meter. The pressure reading can now be
adjusted by using the up or down arrows on the
POV
field
.
A red rectangle highlights the offset to
be adjusted at each stage of the
calibration.
Actual
Chamber
Pressure
Actual Chamber Temperature
Calculated Chamber Pressure (Calculated
from chamber temperature)
The abbreviation definitions are as follows:
POV
-
Pressure offset at Vacuum
POA
-
Pressure offset at Atmosphere
POS
-
Pressure Offset at Sterilization
Temperature
TO
-
Chamber Thermocouple Offset
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...