80
Step-by-step troubleshooting by symptom
PROBLEM
POSSIBLE CAUSE
POSSIBLE SOLUTION
No power
Power cord or main power
issue
Check that the unit is plugged into a
properly grounded outlet and that the
power cord is firmly seated at the rear
of the machine.
Try another outlet. Power unit OFF for
10 seconds and then power ON again.
Check the condition of the line circuit
breaker or fuse.
There is water under the
machine.
Spill over from refilling
External water tank feed
issue
Check that water was not spilled when
refilling the reservoir.
Check that the tube coming from the
external tank (if fitted) is completely
pushed onto the connector.
Run a Vacuum Test. If water drips from
the underside of the unit during the
test, call your SciCan dealer.
Cycle interrupted — NOT
STERILE, Cycle aborted — NOT
STERILE and CYCLE FAULT
messages.
The STOP button was
pressed while the unit was in
operation.
A power outrage or power
fluctuation occurred while
the unit was in operation.
Wait a few minutes and attempt
another cycle before proceeding to the
next solution.
NOTE: STATCLAVE G4 units connected
to the Internet and registered with
SciCan will automatically send Cycle
Fault messages to SciCan’s
international service center.
Excessive steam issuing from
the front of the machine.
Door seal issue
Open and close the door then attempt
another cycle. Check the door seal for
misalignment or damage. Replace the
door seal if required.
If the leak persists, turn the unit
OFF
,
remove the load and contact your
SciCan dealer.
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...