Troubleshooting loading errors and operator complaints
COMPLAINT
POSSIBLE CAUSE
POSSIBLE SOLUTION
Machine will not start, and
touchscreen shows: [H2OX]
Clean water reservoir issue
either water level is too low or
water used is of inadequate
quality.
Press on the icon to confirm
whether it is a water level
problem or a water quality
problem.
If water level is too low: Refill
the reservoir. Refer to the steps
described in section 3.4 Filling
the Reservoir.
If the water quality is
inadequate: You have likely
used water that is not steam-
process distilled or is improperly
distilled.
Empty the reservoir and refill
with steam-process water
containing less than 5 ppm total
dissolved solids (having
conductivity of less than 10 μS /
cm).
If you have a water conductivity
meter, check the quality of the
water before refilling the
reservoir.
To empty the reservoir, see
section XX Shipping the unit.
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...