87
COMPLAINT
POSSIBLE CAUSE
POSSIBLE SOLUTION
Inadequate water quality.
Drain the clean water reservoir
and refill it with high-quality
distilled water.
Organic or inorganic residues on
the instruments.
Instruments must be free of
debris prior to sterilization.
Clean and rinse all instruments
before loading them into the
sterilizer. Disinfectant residues
and solid debris may inhibit
sterilization and damage the
instruments and the STATCLAVE
G4. Lubricated instruments
must be wiped thoroughly, and
any excess lubricant should be
removed before loading.
Contact between instruments
made of different metals.
Separate instruments made of
different metals. See Section 4.3
Unwrapped Instruments for
instructions on arranging
instruments made of different
materials.
Lime residue on the wall of the
sterilization chamber and/or
accessories.
Clean the device and its parts,
as required.
(See Appendix C
“Maintenance”).
Water in the chamber
Cycle interrupted during warm-
up phase.
Remove load. Remove rack.
Use clean cloth or paper towel
to mop up excess water.
Insert rack and load and start a
new cycle.
Unit is using too much water
Unit is overloaded
See Section 4. Loading
Instruments for details on
capacity.
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...