88
COMPLAINT
POSSIBLE CAUSE
POSSIBLE SOLUTION
Door will not close
Item obstruction
Check for instrument or a
cassette that is keeping the
door from closing correctly.
Check the door seal to make
sure it is properly seated. To re-
seat the door seal, see Section
9.13 Replacing the Door Seal.
Door will not close - No
obstructions
Chamber pressure balance
issue.
Leave the door open for 1
minute and try again.
Handle in closed position but
door not showing ‘locked’
The door will lock once a cycle is
selected.
Press a cycle button to initiate
the door lock microswitch.
Touchscreen remains on
WARMING UP CHAMBER screen
Band heaters were not on. From
a cold start, the unit can take
approximately 15 minutes to
warm up. The chamber needs to
be above 5
0°C and the band
heaters need to be at 120°C or
more.
Go to SETTINGS and select
STANDBY. Change unit Stand-by
setting to HIGH.
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...