9
Specifications
Machine Dimensions:
Width: 17.75” / 450 mm
Height: 19-19.5” / 483-495 mm
Depth: 25” / 635mm
Chamber Dimensions:
Diameter: 11”/ 280 mm
Depth: 15”/381 mm
Sterilization Chamber Volume:
6.9 US gal / 26 L
Distilled Water Reservoir Volume:
1 US gal / 4 L
Venturi Reservoir Volume (including condenser):
1.6 US gal /6 L
Weight (without water):
136 lbs/ 61.7kg
Weight (with full reservoirs and full load):
175 lbs/ 79.5kg
Clearance required:
Top: 7” /180 mm
Right Side: 2” / 50 mm
Left Side: 0”/ 0 cm
Front with door open: 16”/ 41 cm
Water quality:
≤6.4ppm / 10 µS (conductivity at 20°C/68°F)
Minimum distilled water required for cycle:
0.26 US gal / 1 L
Minimum Venturi water required for cycle:
0.26 US gal / 1 L
PRV value (pressure relief valve):
Set at 40.6 PSI / 2.8 bar to release pressure in
overpressure situations
Electrical Rating*:
120V, 60Hz, 12 A
208-240V, 60Hz, 12A
*See serial number label for requirements specific to
your unit.
Ethernet Port:
10/100 Base-T
USB Port:
USB 2.0
Current:
AC
Protection Class:
I
Protection:
covered
Ambient Operating Temperature:
41˚F
-
104˚F / 5˚C
-
40˚C
Sound levels:
≤
60 dB
Humidity:
80% maximum
Max. Altitude:
6,562 ft / 2000 m
Safety devices
•
Overheat thermostat: Steam generator overheat thermostat’s
set point 310°C.
•
Overheat thermostat: Band heater overheat thermostat set point 180
°C
.
•
Pressure relief valve: The chamber pressure relief valve is set to 40.6 PSI/ 2.8 bar gauge to release
pressure in an overpressure situation.
•
Pressure or vacuum relief on power failure: The unit will automatically return the unit to atmospheric
pressure when the power is interrupted.
•
Electrical protection: two 15 Amp fuses (high current) and 2 Amp fuse (low current).
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...