90
CF #
Description
Possible causes
Phase of the cycle
CF 28
Chamber pressure went
above threshold
Pressure measurement failure.
Check pressure transducer
Miscalibration.
Blocked exhaust / valve
CF 30
Actual chamber
temperature and
calculated chamber
temperature do not
match
Air leak
Blocked exhaust / valve
Miscalibration
Valve Exhaust Chamber
PRESSURIZING
CF 42
Manual clean water
filling failure
Water filling Valve bad/disconnected
Blocked tubing
Clean Water Level switch stuck OFF
CF 43
Automatic clean water
filling failure
Water filling Valve bad/disconnected
Blocked tubing
External Water filling pump
bad/disconnected
IO board failure
24V power supply failure
Clean Water Level switch stuck OFF
CF 44
Venturi water reservoir
overflow
Water reservoir exhaust blocked
Overflow switch malfunction (stuck ON)
CF 50
Chamber temperature
LESS than
121°C
(for
121°C cycles)
Pressure and/or temperature sensor
miscalibration.
Poor air removal during conditioning
(exhaust blockage).
Not able to generate steam or a leak in
the system.
STERILIZATION
CF 51
Chamber temperature
MORE than
124
°C
(for
121°C cycles)
Pressure and/or temperature sensor
miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed solenoid valve).
STERILIZATION
CF 52
Measured pressure
LESS than calculated
pressure
208 kPa (for
121°C cycle)
Pressure and/or temperature sensor
miscalibration
STERILIZATION
CF 53
Measured pressure
MORE than calculated
pressure
222 kPa (for
121°C cycle)
Pressure and/or temperature sensor
miscalibration.
Unable to depressurize cassette
(blocked exhaust, failed valve).
Poor air removal (partial exhaust
blockage).
STERILIZATION
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...