93
CF #
Description
Possible causes
Phase of the cycle
CF 83
Bottom band heater
temperature sensor
failure
Bottom band heater temperature
sensor failure
Controller board failure
CF 84
Venturi water reservoir
temperature sensor
failure
Venturi water reservoir temperature
sensor failure
Controller board failure
CF 90
Corrupted/not
initialized chamber
thermocouple
calibration
Corrupted/not initialized chamber
thermocouple calibration
CF 91
Corrupted/not
initialized pressure
calibration
Corrupted/not initialized pressure
calibration
CF 92
Controller board
temperature high
Air filter clogged, poor air ventilation
CF 95
Communication Lost
Harness cable between the User
interface board and controller board
damaged or disconnected
CF 98
Communication with the
ADC failed
Controller board damaged
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...