45
Viewing the STATCLAVE G4 from another networked device may be done at high or low resolutions.
NOTE: Select LOW if you want the screens to load faster. Default is HIGH.
Select high or low quality
TECHNICIAN TOOLS
Dealer ID
Enter the number assigned to
the dealer by SciCan
Change Password
Set a new password to limit
access to the service menu
Maintenance
Notification On/Off: Turn the
maintenance message on/off.
Schedule: Select maintenance
intervals by months or cycles.
Reset: Reset the maintenance
counter after performing the
service.
CYCLE SELECTION
Some users prefer their units to display only the most common cycles used. Press on a cycle icon to
remove it from the home screen menu. Press on the icon to reinstate it.
Select the cycle icons you wish to
display on the home screen
Setting up load traceability with USER ID, PIN, and Process Enforced Usage
The Process Enforced function keeps track of who has started and who has removed a load from your
STATCLAVE G4. It does this by prompting users to enter a PIN at the start of a cycle, when they STOP or
CANCEL a cycle, and when they REMOVE a load. Using Process Enforced does not restrict any functions, it
is simply a means of tracking whether a registered user or unregistered user was operating the unit. To use
the Process Enforced feature, you must first assign User IDs and PINs.
Summary of Contents for STATCLAVE G4
Page 5: ...5 Diagram of key systems and components...
Page 6: ...6...
Page 17: ...17 Checking package contents...
Page 18: ...18 Positioning a unit Unit dimensions and operating environment...
Page 20: ...20 Direct to drain connection recommended...
Page 21: ...21 Connecting to a waste bottle...
Page 22: ...22 Connecting to the Internet Connecting to a wired network Connecting to a wireless network...
Page 23: ...23...
Page 25: ...25 Connecting an auxiliary pump...
Page 28: ...28...
Page 30: ...30 Starting and stopping a cycle...
Page 50: ...50 Registering for online access...
Page 51: ...51 First start up...
Page 52: ...52...
Page 58: ...58 Draining the unit for cleaning and shipping...
Page 61: ...61 Replacing the bacteriological filter Replacing the door seal...
Page 75: ...75...
Page 107: ...107 APPENDIX A STATCLAVE G4 plumbing diagram...
Page 108: ...108 APPENDIX B STATCLAVE G4 electrical schematic...