As the cycle is running an intermittent buzzing sound will be apparent as the pump
injects water into the steam generator. A random clicking sound will also be heard as
the exhaust valve opens and closes.
Once the sterilization cycle is complete, a reminder tone will sound and the amber light
will flash as the cycle moves into the Air Drying phase.
The buzzing noise during the air drying stage is the compressor operating. The air
drying phase of the cycle may be interrupted at any time by pressing the
STOP
button.
The display will appear similar to:
[CYCLE YOU SELECTED]
REMOVE CASSETTE
When the automatic 60 minute air drying stage is finished the display reads:
[CYCLE YOU SELECTED]
CYCLE COMPLETE
If a sterilization cycle is successful the reminder tone sounds and the amber light flashes
until the
STOP
button is pressed or the cassette is removed from the unit.
If a message is displayed which gives a CYCLE FAULT code or a NOT STERILE
message, the cassette contents are not sterile. See Section 6. Troubleshooting for
more information.
* Statim 2000 Cycle Time, Temperature, and Model Listing
Page 21
4. Instructions For Use Cont’d.
/
Country
Model no.
Unwrapped
Wrapped
Rubber & Plastics
Temp
Time
Temp
Time
Temp
Time
Czech
01-122328
135˚C
4 min. 135˚C 10 min.
121˚C
20 min.
275˚F
275˚F
250˚F
Canada
(and all
All other
135˚C 3.5 min. 135˚C 10 min.
121˚C
15 min.
other Countries)
models
275˚F
275˚F
250˚F