Page 25
/
4. Instructions For Use Cont’d.
4.9 Statim 5000 — Running a Cycle
See the chart on the end of section 4.9 Statim 5000 — Running a Cycle for a complete
list of cycle time and temperature unit listings in different countries. To operate each
cycle, follow these steps and watch the LCD.
or
14:23 3/11/2001
SELECT A CYCLE
SELECT A CYCLE
PRESS
START
65˚C
WARMING UP
Turn the power switch at the
back of the unit to
ON
.
The LCD Display reads:
(if optional printer is installed)
Press the appropriate cycle
button on the keypad.
The display will read either:
As the button is released the
display appears similar to:
The amber indicator light comes on and the display will appear similar to:
[CYCLE YOU SELECTED]
[CYCLE YOU SELECTED]
110˚C
130˚C
135˚C
212 kPa 3:30
105˚C
CONDITIONING
PRESSURIZING
STERILIZING
VENTING
or
or
UNWRAPPED
135˚C for 3.5min.*
WRAPPED
135˚C for 6 min.*
RUBBER AND PLASTICS
121˚C for 15 min.*
or
HEAVYDUTY UNWRAPPED
135˚C for 6 min.*
As the cycle progresses,
the cycle stage and temperature
of the chamber are displayed.
During sterilization, the pressure
of the chamber and the time remaining
for that stage is also displayed. During
the air drying phase, the time remaining
is displayed.
* See the chart on the end
of section 4.9 Statim 5000-
Running a Cycle for a
complete list of cycle time
and temperature unit listings
in different countries.