Page 111
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
5. STAT
IM 5000S Calibration
STAT
IM
5000/5000S/5000 G4 Service Guide
7. Turn ON the unit while keeping Unwrapped and Wrapped button pressed to enter STAT
IM
Service Mode
.
8. This
Service Mode
is password protected, enter password to continue (default password is:
Unwrapped, Wrapped, R&P and Stop keys pressed in that order).
Keypad functions (Unit LCD or Control Box) at this time will be:
• unwrapped key:
Select next item in the menu.
• Wrapped key:
Select previous item in the menu.
• Rubber and Plastics key:
Enter current selection
• Toggle
using keypad through the menu selection to reach Calibration option and press
R&P key.
The display should be similar to the diagram below.
Where
• PPP.P = actual chamber pressure in kPa.
• PO = chamber pressure hexadecimal offset value
• TTT.T = theoretical pressure calculated from the actual chamber temperature.
• CO = chamber temperature hexadecimal offset value
• CCC.C = actual chamber temperature in °C
IMPORTANT NOTES:
• Remember – the steam generator thermocouple does NOT require calibration.
• The VREF does not need adjustment.
PPP.P PO TTT.T
CO
CCC.C