Page 115
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
5. STAT
IM 5000S Calibration
STAT
IM
5000/5000S/5000 G4 Service Guide
7. Turn ON the unit and access the service menu in the following manner:
8. This Service Mode is PIN protected, enter password to continue (default password is: 7919).
9. From the service menu, select Calibration.
IMPORTANT NOTES:
• Remember – the steam generator thermocouple does NOT require calibration.
• The VREF does not need adjustment.
1. Chamber Thermocouple and Pressure Transducer Calibration
To calibrate the chamber thermocouple and pressure transducer:
1. To start the calibration cycle, select Calibration in the Calibration submenu and calibration will
start automatically