Page 164
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
8. Water Pumps, Reservoir, and Compressor
Water Pump Recovery Repair Procedure
Pump performance may be compromised because of foreign material caught inside the assembly,
blockage of an orifice or mechanical wear of internal parts. Before replacing a pump that is
performing poorly, attempt to recover the unit by cleaning the filter(s) and/or resizing the pump tube.
Performing Filter Maintenance
Early STAT
IM 5000 / 5000S units fitted with SciCan pumps have a coarse mesh filter in the inlet
fitting and a fine mesh filter in the outlet elbow fitting located internally in the pump.
In 2007, SciCan removed these two filters due to the implementation of the water reservoir filter in
2006 and subsequent introduction of the in-line filter in 2007.
The in-line filter was added between the water
reservoir and the water pump inlet to address the
issue of customers not using the reservoir filter.
The purpose of this in-line filter is to stop debris
from entering the STAT
IM plumbing system when
the reservoir filter is not used. The in-line filter kit
is available as part # 01-106637S and includes the
filter and all necessary tubing. This filter should be
replaced every 2000 cycles or every two years.
Filter in the inlet fitting
prior to Jun ’07
In-line filter
Filter on the outlet fitting prior to Jun ‘07
Figure 5
Figure 4
STAT
IM
5000/5000S/5000 G4 Service Guide