Page 34
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
2. Tools, Maintenance Schedules, Procedures and Testing
Routine Maintenance Schedule
and Procedures
To maintain the STAT
im
in good order between annual services, follow these guidelines:
DAIlY
Water Reservoir
• Replace water as needed.
• For opthalmic use, drain at the end of every workday, leave empty, and refill at the start of
the next workday.
Water Bottle
• Empty the waste bottle every time you refill the reservoir.
• Fill with water, up to MIN line marking. You may also add some chlorine-free disinfectant.
WeeklY
Cassette
• Wash the interior of the cassette with dishwashing soap or a mild detergent that does not
contain chlorine.
• Scrub the inside with a cleaning pad designed for use with Teflon™-coated surfaces.
• After removing all traces of the detergent, treat interior surfaces of the cassette with the
STAT-DRI™ Plus drying agent to enhance the drying process. Order more STAT-DRI™ Plus
from SciCan quoting 2OZPLUS, 8OZPLUST, or 32OZPLUS. Please note that STAT-DRI
drying agent is not be to used with the U.S. G4 models (G4-121101 and G4-201103).
Biological and/or Air Filter
• Check the filter for dirt and moisture. Replace if dirty. Call for service if wet.
Water Reservoir Filter
• Check the water reservoir filter every week and clean if necessary. Replace only if
necessary.
eveRY 6 MoNTHS
Cassette Seal
• Replace every 500 cycles or six months (whichever is first), or whenever necessary.
Biological and/or Air Filter
• Replace every 500 cycles or six months (whichever is first).
STAT
IM
5000/5000S/5000 G4 Service Guide