Page 36
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
2. Tools, Maintenance Schedules, Procedures and Testing
STAT-DRI should be applied every 10 cycles, and after every cassette cleaning.
STAT-DRI is available from your SciCan Dealer. Please note that STAT-DRI drying agent is not be
to used with the U.S. G4 models (G4-121101 and G4-201103).
Cleaning the Water
Reservoir Filter
(where fitted)
This water reservoir filter should
be cleaned at least once a week
or when required. The filter can
easily be removed and cleaned
by placing the filter upside down
under running water to wash away
the particles until clean, and then
placed back into the reservoir
opening. Replacement water
reservoir filters are available from
you SciCan dealer.
Cleaning the Water Reservoir
Check the reservoir for dirt or particles. The reservoir may be cleaned by draining followed by
cleaning and rinsing with steam process distilled water ONLY. Use of chemicals or cleaning
agents is not recommended and could cause damage to the unit.
Changing the air filter and bacteria retentive filter (where fitted)
Never re-use an old filter. Never run the unit without a clean filter in place.
After each sterilization cycle, the compressor forces air through the cassette to rid it of steam and
to dry the instruments. The air is drawn into the compressor through one or two filters located at
the back of the unit, depending upon which unit is being serviced.
Note:
that some units may not have both or either filters fitted depending on age, unit type,
compressor type, or if a compressor is fitted.
Filters on the STAT
IM 2000/2000S/2000 G4
The STAT
IM 2000 type units draw air into the compressor (where fitted) through a circular foam air
filter located behind a cover plate on the rear of the compressor. This plate can be accessed from
the rear of the unit with the cover still attached. The air is then directed to the steam generator and
cassette, by the compressor, through a bacteria-retentive air filter (where fitted) which is held in a
bracket attached to the rear cover of the unit. If both filters are present, always change both filters
at the same time.
STAT
IM
5000/5000S/5000 G4 Service Guide
Figure 21