Page 48
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
3. Troubleshooting Cycle Faults
3. Turn the STAT
im
ON. The unit is now in calibrating mode.
4. For version 2 and 3 software without selecting a cycle, press the START button. For other
versions of software press the “Wrapped” button.
5. The printer will print the last complete or incomplete cycle. This information can be useful in
diagnosing intermittent problems when the cycle fault code has not been recorded.
For units with revision 7.x controllers
(with service menu), including G4
All revision units fitted with revision 7.x controllers have a service menu which is designed to:
• Provide information on the setup of the machine.
• Allow the changing or resetting of various machine parameters.
• Operate various aspects of the unit with the cover on as opposed to using a control box.
Certain information provided in this menu may be used to assist in diagnostics.
using the service menu on the STATIM Classic
(Rev. 7 controller board)
The service menu can be displayed on the LCD screen of any STAT
im
Classic with a Rev. 7 board
by leaving the cover on, or using the LCD screen of the control box if the cover is removed and
detached from the unit.
• To enter the Service Menu, power ON the STAT
im
while holding down the “Unwrapped”
and “Wrapped” keys.
• You will then be prompted to enter the service password to allow you to enter the
service menu.
• The default password is entered by pressing the “Unwrapped”, “Wrapped”, “Rubber and
Plastics” and “Stop” keys in that order.
• The service technician can change this password. In case the changed password is
lost or has been changed by another technician, a backdoor password can be used
(which can not be changed) by pressing the “Unwrapped”, “Wrapped”, “Unwrapped”,
“Wrapped” keys in this order.
STAT
IM
5000/5000S/5000 G4 Service Guide