Page 59
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
3. Troubleshooting Cycle Faults
STAT
IM
5000/5000S/5000 G4 Service Guide
>Press. Sensor Type – (S class units oNlY)
Screen display when in Press. Sensor Type sub menu.
This sub menu is for information only.
All S class STAT
im
units are fitted with a pressure transducer that actively measures the pressure
in the chamber. Two types of transducer were fitted (
60
p.s.i. and
68
p.s.i.) depending on the age
and controller type of the unit.
Both transducers are visually the same but they are
NOT
interchangeable between units as each
transducer requires type specific software for the unit to operate correctly.
Notes on how to identify the different types can be found in Chapter 9 of the service manual, but
units with revision 7 controllers (including upgraded older units with the correct software) will show
the pressure transducer type in this sub menu.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
>validation offset – (Non S class units oNlY)
Screen display when in Validation Offset sub menu.
This sub menu is for information only.
## = unit validation thermocouple offset value.
This allows the validation thermocouple offset value to be read without entering calibration mode.
No changes can be made to the offset value in this sub menu.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
>voltage Calibration.
Screen display when in Validation Offset sub menu.
This sub menu is for STAT
im
7000 units ONLY and is not used on 2000/5000 units.
1. To return to the service menu, press the RUBBER & PLASTICS button.
2. To return to the operational menu, press the STOP key.
>Press. Sensor Type
68PSI
>Validation Offset
##
>Validation Calibration
V = VVV VCal. =CCC