Page 62
STAT
IM 2000/2000S
Service Guide
96-106775 Rev 5.0
3. Troubleshooting Cycle Faults
STAT
IM
5000/5000S/5000 G4 Service Guide
To activate NVRAM backup, proceed as follows:
1. The UNWRAPPED or WRAPPED keys will both switch the screen display to “
YES
”.
2. The RUBBER & PLASTICS key will action your selection and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without changes and return the unit to the operational menu.
>Restore NvRAM
Screen display when in Restore NVRAM sub menu.
This function will allow the stored (backed up) unit settings to be restored if operational settings
are lost for any reason.
The default setting on all STAT
im
units is
NO
.
The alternative is
YES
.
To activate NVRAM restoration, proceed as follows:
1. The UNWRAPPED or WRAPPED keys will both switch the screen display to “
YES
”.
2. The RUBBER & PLASTICS key will action your selection and return the unit to the service menu
condition.
3. Pressing the STOP key will exit without changes and return the unit to the operational menu.
>Save and exit
Allows any changes made in the sub menus (where the STOP button does not action save) to be
saved and the unit will then exit to the operational menu.
To activate save and exit, press the RUBBER & PLASTICS button.
>exit
Allows you to exit the service menu WITHOUT saving changes other than those changes saved by
activating the STOP button.
To activate exit, press the RUBBER & PLASTICS button.
>Water Pump Type
Screen display when in Press. Sensor Type sub menu.
This sub menu is for information only.
>Restore NVRAM
No
>Water Pump Type
Ulka