6) Focusing, diopter, interpupillary adjustments:

Place a specimen onto the stage. Loose the body locking thumb screw and hold the microscope head and move 

the body up and down to fix it at the estimated working distance. Rotate the zooming knob while looking 

through the right eyepiece until you see the image. Using the focusing handles to get the sharpest image of the 

specimen. Then look through the left eyepiece with your left eyes and turn the diopter adjustment ring until you 

get an image as sharp as the right side. Make this adjustment without moving the focusing handle. Then grasp 

the right and left prism housing and move them closer or father apart in order to match your pupil distance. Ad-

justment is proper when the field of view becomes comfortable and presents a full single field.

7) Trinocular zooming head:

When the photo switch pressed in, the microscope is set for binocular users. When it is pulled out, the photo 

tube is in working condition. Adapters for cameras, digital cameras, CCD are optional if necessary.
8) Use of auxiliary objectives:

Auxiliary objectives of 0.75X, 1.5X, 2X can be screwed in directly at the tip of the objective cases. 

Because the working distance of the 0.5X objective is long, long pole should be used before it is used. 

(Please refer to optical index III).

REPLACEMENT OF LAMP AND FUSE:

  Warning: Disconnect the power line before you change the lamp or fuse.

• Replacement of the incident lamp:

 

Loosen the fixing screw and take off the lamp housing. Replace the bulb with a same new bulb. Place the 

lamp housing back and fix it with the same screw.

• Replacement of the transmitted lamp:

 

Loosen the fixing screw of the glass stage and take off the glass. Take off the broken bulb through the stage 

hole and install a new bulb.

• Replacement of the fuse:

 

The fuse case is located at the back side of the base. Unscrew the fuse case cover and put in a new one.

MAINTENANCE AND GENERAL CARE OF YOUR MICROSCOPE:

1. Microscope is a delicate precision instrument and it may be damaged by dropping and hitting.

2. Do not keep microscope under sun. It should be kept in a dry and clean environment and avoid heat and 

strong tremor.

3. To obtain clear image, do not touch lenses with your finger.

4. All lens surface should be kept clean. If the lens get dusty blow off the dust with a rubber syringe. If neces

-

sary clean the lenses with a lint free cloth dipped in aether.

5. Do not use any organic material to clean the microscope surface, especially the plastic surface. It should be 

cleaned by neutral detergent.

6. Because the assembly of all parts has been done by skilled optical craftsmen at the factory, you should never 

attempt disassembly.

7. Apply a little bit grease regularly to the mechanical parts.

8. When not in use always cover the microscope with the dust cover and place it in a cool and dry place.,