background image

                                                                                     

  
 

 

                

 

Standard Warranty 

Applies Only Within the Continental United States 

 

SCIENTEMP WARRANTS TO THE ORIGINAL PURCHASER 
FIRST 18 MONTHS – 

The cabinet and all of its parts shall be free of defects in material and workmanship under 

normal use and service for a period of 18 months from the date the unit has been shipped from our facility. 
Scientemp’s sole obligation under this warranty shall be limited, at its option, to either repairing or replacing any 
part of the cabinet determined by an authorized service agent to be defective. Scientemp reserves the right to repair 
the freezer at our facilities. 
 

THE SCIENTEMP STANDARD WARRANTY DOES NOT COVER 
 
TRANSPORTATION COSTS – 

Scientemp shall not be responsible for transportation or incidental costs incurred 

in connection with the repair or replacement of a cabinet or any of its parts. 
 

ABUSE, MISUSE, ACCIDENTS – 

Scientemp shall not be responsible for parts or assemblies which upon 

inspection are determined by an authorized Scientemp Service Agent to have been subjected to misuse, neglect, 
alteration, accident, abuse, damage during transit or delivery, or by fire or flood.  
 

CONSEQUENTIAL DAMAGES – IN NO EVENT SHALL SCIENTEMP CORP BE LIABLE FOR 
SPECIAL, INDIRECT OR CONSEQUENTIAL DAMAGES, INCLUDING WITHOUT LIMITATION, 
PRODUCT LOSS OR PRODUCT SPOILAGE CLAIMS, NOR FOR ANY DELAY IN THE 
PERFORMANCE OF THIS WARRANTY DUE TO CAUSES BEYOND ITS CONTROL. 
 
GENERAL 

The standard warranty and any service contract related to the STANDARD WARRANTY shall apply only to the 
products sold and used within the boundaries of the Continental United States. 
 
Users may file warranty claims either directly with Scientemp Corp, 3565 S. Adrian Hwy, Adrian, MI 49221, or 
with the seller from whom the cabinet was purchased. All claims must be supported by information concerning the 
alleged defect and specifically identified by the Serial Number of the cabinet. 
 

THERE ARE NO OTHER WARRANTIES EXPRESS, IMPLIED, OR STATUTORY, EXCEPT THIS 
WARRANTY, WHICH IS IN LIEU OF ALL OTHER WARRANTIES INCLUDING TO THE EXTENT 
PERMITTED BY LAW, ANY IMPLIED WARRANTY OR MERCHANTABILITY OR FITNESS FOR A 
PARTICULAR PURPOSE. 
 

DO NOT DRILL HOLES IN CABINET 

Refrigeration tubing and wiring is routed through the cabinet walls. Leaks, wet insulation or electrical problems 

caused by drilling holes are not covered by warranty. 

 

 

                   3565 S. Adrian Hwy 

 Adrian, MI 49221 

800-968-2653 

517-263-6020 

Fax  517-263-5492 

Summary of Contents for 43-1.7A

Page 1: ...ure When the temperature has been reached turn the alarm switch to the On position Cabinet should not be loaded with product until the cabinet has operated for 24 hours at the desired set point temperature Maintenance tips for your Scientemp Freezer Regularly remove any frost build up by using a plastic wooden scraper Check and replace batteries on Alarm Signal System and Chart Recorder if equippe...

Page 2: ...517 263 5492 Website www scientemp com OPERATING AND INSTALLATION MANUAL MODEL 43 1 7A This cabinet has passed the QUALITY CONTROL INSPECTION and meets the high standards at Scientemp Corp This inspection includes the complete refrigeration system cabinet construction and finish ...

Page 3: ...lling using or maintaining this product please be sure to read this manual carefully Failure to follow these instructions may cause this product to malfunction which could result in injury or damage Below are important safety precautions that apply to this product Use this product only in the manner described in the product literature and in this manual Before using it verify that this product is ...

Page 4: ...n this is desired In the event a temperature spread of more or less than factory set is desired refer to your digital temperature control manual NOTE It is recommended to keep the freezer operating continually rather than turning the freezer off and on 5 ELECTRICAL CONNECTIONS 5 1 Wiring This cabinet is equipped with a three prong grounding plug for your protection against shock hazards The cabine...

Page 5: ...to the material stored in the freezer Most of the frost and ice collects near the top edge and is easily removed 2 In the case where the freezer is not loaded frost and ice may be removed by disconnecting the power until the frost and ice thaw After defrosting dry the freezer compartment interior and plug freezer back in 8 ALARM SIGNAL SYSTEM SCIENTEMP Alarm signal system is a dependable compact u...

Page 6: ...troller 4 Remove the controller from the instrument panel 5 Insert new controller into instrument panel reverse procedures When rewiring use enclosed wiring diagram 11 CHANGING THERMOCOUPLE T C 1 Disconnect power and let freezer warm up 2 Remove stainless steel corner slips from front two corners 3 Remove white plastic breaker by pulling up and out on edge next to inner liner 4 Lift insulation and...

Page 7: ...play Setpoint value or program selection Green LED Setpoint 1 output indicator Upper Red LED Setpoint 2 output indicator Lower Red LED Setpoint 3 output indicator Green Display Process variable or Function Option ADJUSTMENTS To enter or exit program mode Press together for 3 seconds To scroll through functions Press or To change levels or options Press together or together To view setpoint units P...

Page 8: ... or by fire or flood CONSEQUENTIAL DAMAGES IN NO EVENT SHALL SCIENTEMP CORP BE LIABLE FOR SPECIAL INDIRECT OR CONSEQUENTIAL DAMAGES INCLUDING WITHOUT LIMITATION PRODUCT LOSS OR PRODUCT SPOILAGE CLAIMS NOR FOR ANY DELAY IN THE PERFORMANCE OF THIS WARRANTY DUE TO CAUSES BEYOND ITS CONTROL GENERAL The standard warranty and any service contract related to the STANDARD WARRANTY shall apply only to the ...

Page 9: ......

Page 10: ...17B 3 1 14 4 5 6 10 9 14 15 17A 18 13 7 19 13 8 9 16 2 12 ...

Page 11: ...Width 1095 1 7 W 10 Stainless Steel Corner Cap 1098 1 7 11 Replacement Key Not Pictured 1031 12 Metal Sub Breaker Clip Call 13 Sub Breaker Strip Length Call 14 Sub Breaker Strip Width Call 15 Keyed Hasp Lock Assembly 1027 4 16 Cabinet Shell Call 17A Leg Assembly Left Call 17B Leg Assembly Right Call 18 Standard Non Locking Casters 1030 2 18 Locking Casters Optional Not Shown 1030 2L 19 Interior Ta...