4. CHECK OUT YOUR NEW FREEZER 

4.1 Lid Seal 

The gasket on the door has a magnetic strip to help door seal. 
Be  sure  to  keep  the  gasket  clean  and  free  of  debris.  Clean 
any  frost  that  may  build  up.  If  the  gasket  does  not  seal 
properly,  the  freezer  may  experience  problems  with 
maintaining temperature and excessive frost build up. 

 

4.2 Interior Compartment Temperature 

The  temperature  is  controlled  by  a  CAL  9500  digital 
temperature controller. When you first power up the freezer,  
the high stage compressor and fan will start and operate for 3 
to  6  minutes  before  the  low  stage  compressor  starts.    Pull 
down  to  minimum  temperature  will  take  from  3  to  6  hours, 
depending on ambient temperature and size of cabinet.  The 
high stage compressor runs 100% of the time.  The low stage 
compressor will cycle on and off to maintain the temperature 
of  the  controller  setting.    Set  the  control  to  the  desired  set 
point. To view the set point temperature, press the * and the 
UP  arrow  keys  together.  To  decrease  the  set  point 
temperature,  press  the  *  and  the  DOWN  arrows  keys 
together. 

Freezers  are  shipped  from  the  factory  with  the  indicating 
digital control programmed for proper operation. All but the 
temperature  settings  may  be  adjusted.  The  temperature 
sensor is a type 

“T”

 thermocouple located on the side wall of 

the inner-liner. 

4.3

 

Alarm  

The  Alarm  should  be  kept  in  the  “OFF”  position  until  the 
cabinet  temperature  has  reached  the  appropriate  setting.  
When  the  temperature  has  been  reached,  the  temperature 
alarm  should  be  switched  to  the  “ON”  position  and  is  now 
ready to operate. 
 
The Alarm is equipped with a time delay relay to prevent the 
alarm  from  going  off  every  time  the  door  is  opened.  This 
relay  can  be  adjusted  from  1  to  10  minutes.  This  delay  is 
factory set for 10 minutes. 
 
To adjust time delay relay: 

a)

 

Remove  screw  located  on  right  side,  middle  of 
instrument  panel  with  temperature  controller. 
Instrument panel will hinge open. Time delay has a 
dial and is located on lower right corner of electric 
box. 

b)

 

If you have a chart recorder:  Open glass door to get 
to  the  black  instrument  panel  with  chart  recorder. 
Remove  screw  located  on  right  side,  middle  of 
panel. Instrument panel will hinge open. Time delay 
has  a  dial  and  is  located  on  lower  right  corner  of 
electric box. 

  

NOTE:  It  is  recommended  to  keep  the  freezer  operating 
continually rather than turning the freezer off and on. 

4.4

 

Chart Recorder (Optional ) 

If your freezer has a chart recorder installed. On initial start 
up you must replace the round paper chart, connect the 9 volt 
battery and remove the black rubber cap on pen.  

WARNING: 

Do not lift the pen arm in order to slide a new 

chart under the pen or to remove the black rubber cap.  

To replace the paper chart: Press and hold the “change chart” 
button (#3) for approximately one second until pen begins to 
move to the left of the chart and then release the button. Wait 
until the pen has moved completely off the chart paper. To 
remove the chart paper, unscrew (counter-clockwise) the 
knob at the center of the chart paper. Remove old paper and 
position new one so that the correct time line coincides with 
the time line groove on the chart plate. Re-attach the knob, 
screw securely (by hand). (Do not over tighten. Over 
tightening can damage the gears in the chart motor gearbox 
assembly.) Press and hold change chart button (#3) again 
until pen begins to move back and then release button. 

To remove the black rubber cap at start up:  While you have 
arm moved off the paper for changing the chart, remove the 
rubber cap.

 

 

5 ELECTRICAL CONNECTIONS 

5.1 Wiring 

 

 

This cabinet is equipped with a three-prong (grounding) plug 
for your protection against shock hazards.  The cabinet 
should be plugged directly into a properly grounded three-
prong receptacle.  

Where a two-prong wall receptacle is encountered, it must be 
replaced in accordance with the National Electronic Code 
and local codes and ordinances. A licensed electrician must 
do the work. 

The electrical outlet should not be controlled by a wall 
switch, which might be turned off accidentally. 

 

*WARNING: Do NOT under any circumstances cut or 
remove the round grounding prong from the cabinet 
plug. 

 
 

*CAUTION

: For personal safety and trouble-free operation, 

this cabinet must be properly grounded before it is used.  
Failure to ground

 

the

 

equipment may cause personal injury 

or damage to the equipment.  Always conform to the 

Summary of Contents for 45-01A

Page 1: ...ase button Provide at least 3 of space around the freezer for ventilation Avoid placing the freezer in locations exposed to direct sunlight heat registers or any other heat source Alarm On Off switch should remain in the Off position until the freezer reaches desired set point temperature When the temperature has been reached turn the alarm switch to the On position Cabinet should not be loaded wi...

Page 2: ...es not have a Chart Recorder Open up to the compartment behind and Next gently separate and slide the right grill panel off you will see two more screws DO NOT remove the left side 2 3 Remove these screws 4 Slide this part completely off to the right 5 After the grill is removed you will see the black condenser Being careful not to bend the steel fins vacuum vertically up and down ...

Page 3: ... 517 263 5492 Website www scientemp com OPERATING AND INSTALLATION MANUAL MODEL 45 01A This cabinet has passed the QUALITY CONTROL INSPECTION and meets the high standards at Scientemp Corp This inspection includes the complete refrigeration system cabinet construction and finish ...

Page 4: ...his product only in the manner described in the product literature and in this manual Before using it verify that this product is suitable for its intended use Do not modify system components The cabinet must be properly grounded in conformity with national and local electrical codes Never connect the unit to overloaded power sources Disconnect the unit from all power sources before cleaning troub...

Page 5: ...anel with chart recorder Remove screw located on right side middle of panel Instrument panel will hinge open Time delay has a dial and is located on lower right corner of electric box NOTE It is recommended to keep the freezer operating continually rather than turning the freezer off and on 4 4 Chart Recorder Optional If your freezer has a chart recorder installed On initial start up you must repl...

Page 6: ...rmer areas of your body If possible immerse your hands in water that is slightly above normal body temperature about 100 F or 37 8 C or which feels warm to someone else until your normal color returns Do not use direct heat Seek medical attention as quickly as you can 7 REFRIGERATION COMPARTMENT The refrigeration system of a Scientemp freezer is hermetically sealed It requires no maintenance Howev...

Page 7: ...play Setpoint value or program selection Green LED Setpoint 1 output indicator Upper Red LED Setpoint 2 output indicator Lower Red LED Setpoint 3 output indicator Green Display Process variable or Function Option ADJUSTMENTS To enter or exit program mode Press together for 3 seconds To scroll through functions Press or To change levels or options Press together or together To view setpoint units P...

Page 8: ...arameter if calibrating unit value may vary ChEk off rEAd Var tECh CtA Ver 953 rSET none Level 2 SP1 P 0 hAnd off PL 1 100 PL 2 100 SP2 A Dv hi SP2 b none di SP 0 1 hi SC 100 Lo SC 100 in Pt tc t unit C Level 1 tune off band 0 3 int t off dEr t off dac 0 5 cyc t on of ofSt 0 0 SP Lk off SPrr 0 SPrn off SoAk Set 2 6 alarm deviation bnd 2 0 1 Cyc 2 on of ...

Page 9: ... or by fire or flood CONSEQUENTIAL DAMAGES IN NO EVENT SHALL SCIENTEMP CORP BE LIABLE FOR SPECIAL INDIRECT OR CONSEQUENTIAL DAMAGES INCLUDING WITHOUT LIMITATION PRODUCT LOSS OR PRODUCT SPOILAGE CLAIMS NOR FOR ANY DELAY IN THE PERFORMANCE OF THIS WARRANTY DUE TO CAUSES BEYOND ITS CONTROL GENERAL The standard warranty and any service contract related to the STANDARD WARRANTY shall apply only to the ...

Page 10: ...CK WHITE CAL 9500 CONTROLLER 6 5 4 3 2 1 N L INPUT TC COMPRESSOR RECORDER RED POWER RELAY RED WHITE BLACK WHITE WHITE RED BLACK BLUE RED BLACK BLACK L1 BLACK BLACK BLACK WHITE ORANGE ORANGE BLACK WHITE RED RED WHITE WHITE L2 BLACK WHITE RTD RED RED RED RED BLACK RED RED RED 8 7 45 01 WIRING DIAGRAM CAL 9500 WHITE FAN ...