11
Examples:
The actual measurement data of the third-party sensor is 37.00℃, and the value before CLP1
correction is 37.05℃, then the COP1 deviation value should be entered -0.05
(deviation value=actual measurement data - pre-correction value),
press button (B), to save the deviation value.
CLP1
Before correction
, COP1
Deviation
Note: Make sure the calibration value is stable and reliable, and then press the button (B) ,to save the
deviation value.
Show device serial number:
To show the 1st 4 Digi of the serial number(High Serial number——HiSn)
Press button (B),enter the menu to show another 4 Digi of the serial number.
Press button (A),to exit.
Press button (C) or (D) to show the 1st 4 Digi of the serial number.
To show another 4 Digi of the serial number(Low Serial number——LoSn)
Press button (B),enter the next menu.
Press button (A),to exit.
Press button (C) or (D) to show another 4 Digi of the serial number.
Summary of Contents for SCIP5-Mini
Page 1: ...Operation Manual Ver 1 7...