background image

SK-O(L)330- Pro  Orbital and Linear Digital Shaker

SK-O(L)180- Pro  Orbital and Linear Digital Shaker

Linear Digital shaker(7.5Kg)

Orbital Digital shaker(7.5Kg)

Linear Digital shaker(2.5Kg)

Orbital Digital shaker(2.5Kg)

SK-L330-Pro

SK-O330-Pro

SK-L180-Pro

SK-O180-Pro

Please read the User Manual and the related Video of this 

instrument on our website carefully before use, and follow 

all operating and safety instructions!

The website www.scilogex.com

User Manual

Cat No.12300140

Summary of Contents for SK-L180-Pro

Page 1: ...igital shaker 7 5Kg Linear Digital shaker 2 5Kg Orbital Digital shaker 2 5Kg SK L330 Pro SK O330 Pro SK L180 Pro SK O180 Pro Please read the User Manual and the related Video of this instrument on our...

Page 2: ...4 Trial Run 3 5 Control and Set 4 5 1 Control 4 5 2 Set 5 5 2 1 Set Time 5 5 2 2 Setting Speed 5 6 Operating Modes 6 6 1 Mode A 6 6 2 Mode B 7 6 3 Switching the Mode 7 6 4 Switch the shaking directio...

Page 3: ...rmation Serial Number on the rear panel Description of problem i e hardware or software Methods and procedures adopted to resolve the problems Your contact information Warranty You have purchased a Sc...

Page 4: ...s substances or under water Observe the vessel on the shaking table for medium splashing out of liquid when setting the motor speed Reduce the motor speed if the instrument is not running smoothly Acc...

Page 5: ...r for technical support Note If there is any apparent damage to the system please do not plug it into the power line 3 2 Listing of Items The packing includes the following items Items Qty Main unit 1...

Page 6: ...wnward with the wrench until the instrument is standing securely on the surface Figure 1 5 Control and Set 5 1 Control Shaking table Time knob LCD time LCD speed Figure 2 Items Descriptions Speed knob...

Page 7: ...ion If continuous mode is selected the shake can continue its shaking function for any amount of time with the previously set speed If timed mode is selected the shaker can run in the set time After t...

Page 8: ...the upper speed limit rpm which can be set by pressing the speed knob and turning at the same time Figure 3 The LCD displays run modes StA A or B for about 2 seconds Figure 4 The set or stored speed v...

Page 9: ...splays the upper speed limit rpm which cannot be changed Figure 7 The LCD displays run modes StA for about 2 seconds Figure 8 The set speed value is shown in the left display and the set time value is...

Page 10: ...curely in the middle of the shaking table no matter one or multiple shaking vessels are placed 8 Faults When switched ON the instrument doesn t work Check whether the power cable fitted well Check whe...

Page 11: ...ase request information from manufacturer or your supplier Before using another than the recommended method for cleaning or decontamination the user must ascertain that this method does not destroy th...

Page 12: ...brushless motor Motor rating input W 28 Motor rating output W 15 Permissible ON time 100 Speed range rpm 100 350 SK L330 Pro 100 500 SK O330 Pro 100 350 SK L180 Pro 100 800 SK O180 Pro Speed display L...

Page 13: ...0V 50Hz 60Hz 832010039999 SK O180 Pro Orbital Digital Shaker 2 5Kg No Attachment Euro plug 110V 220V 50Hz 60Hz Cat No Descriptions 832010049999 SK O180 Pro Orbital Digital Shaker 2 5Kg No Attachment U...

Page 14: ...ker 7 5 Kg No Attachment USA plug 110V 220V 50Hz 60Hz Accessories Accessories Figure Cat No and Designation Operation mode 18900027 SK330 1 SK 330 1 Universal Attachment 18900040 SK330 3 SK330 3 Lengt...

Page 15: ...lask Volume 50 ml 18900031 SK330 2 3 Fixing Clips for Flask Volume 100 ml 18900032 SK330 2 4 Fixing Clips for F l a s k Vo l u m e 200 250 ml 18900033 SK330 2 5 Fixing Clips for Flask Volume 500 ml 18...

Reviews: