32
CHAPTER 4: OPERATION
087-0038 REV L
SCOTT SAFETY
•
Over-Range Conditions – If a sensor within the device is exposed to a target gas
concentration that exceeds the sensor’s range, the device activates its visual,
audible, and vibratory alarms and di++ at the sensor’s location on the
LCD. Alarm conditions clear when the gas concentration has subsided.
Warning:
If an over-range condition is encountered, perform a bump test as soon as
safely possible to ensure device functionality. If the device fails the Bump Test,
perform a full calibration procedure. Failure to do so could result in inaccurate
monitoring by the device and cause death or injury.
Warning:
If using the device with a failed sensor, the failed sensor reading will be
wrong, cannot be relied upon and the device should be replaced at the earliest
possible moment.
•
Low Flow Pump Warning – Displays when airflow through the pump is
diminished or restricted and could compromise accurate readings. Device shuts
down the pump and the device goes into audible, visual, and vibratory alarm while
the PUMP icon flashes. After the cause of the low pump flow is cleared, reset the
pump by pressing and releasing the Operation Button (Left or Right).
•
Sensor Failures – FAL indicates each sensor that fails to Zero or Span Calibrate.
Should this occur, recalibrate the device. If a sensor continues to fail, replace the
sensor and recalibrate or contact an authorized service center.
Summary of Contents for Protege
Page 1: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Page 2: ......
Page 3: ...Protege User Guide PROTEGE PORTABLE GAS MONITOR 087 0038 Rev L ...
Page 10: ...X LIST OF TABLES 087 0038 REV L SCOTT SAFETY ...
Page 44: ...20 CHAPTER 2 INSTALLATION 087 0038 REV L SCOTT SAFETY ...
Page 60: ...36 CHAPTER 4 OPERATION 087 0038 REV L SCOTT SAFETY ...
Page 76: ...52 CHAPTER 5 MAINTENANCE 087 0038 REV L SCOTT SAFETY ...
Page 80: ...56 APPENDIX A SPECIFICATIONS 087 0038 REV L SCOTT SAFETY ...
Page 86: ...62 APPENDIX B QUICK REFERENCE 087 0038 REV L SCOTT SAFETY ...
Page 90: ...66 APPENDIX C SENSOR INFORMATION 087 0038 REV L SCOTT SAFETY ...
Page 102: ...78 INDEX 087 0038 REV L SCOTT SAFETY ...
Page 103: ......