Modular Assembly and Disassembly
1. Position the sensor head so that the dowel pins
align with the holes.
2. With the initial head midway through, reposition it so
that it is parallel to the body. Press the sensor head
firmly against the body until both surfaces are flat
with each other.
3. Insert and screw in the antenna making sure that it’s
not too tight or too loose. Do not use any tools other
than your hand. NOTE: Antenna is not applicable
for the RF Signal Sensor.
Error Message
Reason for Error
ERROR: Sensor is
disconnected. Click here
to turn device off.
The meter is turned
on without the sensor
module.
ERROR: Card was
removed. Data may be
lost and card could be
damaged. Click here to
close the window.
The storage card
(microSD) is removed
while the meter is
powered on.
WARNING: No memory
card. You can add it at
any time. Click here to
close the window.
The meter is turned on
with no storage card in
place.
PREPARING CARD FOR
RECORDING!!! DO NOT
REMOVE!!! PLEASE
WAIT.
A new card is inserted or
when a card is empty.
WARNING: This card
contains the data.
Remove the card or click
here to erase the data.
Inserting a card that has
contents other than the
EM Eye Meter files.
ERROR: Invalid memory
card. Format or replace
it. Click here to close the
window.
The card is not formatted
to FAT32. Formatting is
done using the PC.
ERROR!!! STACK
OVERFLOWED. Click
here to turn device off.
This is a fatal error. If
it appears even after
resetting, please contact
SCS Customer Service
.
Reset Button
If the EM Eye Meter locks up or the display freezes,
press the reset button to restart the meter. The reset
button is in a small hole found near the headphone jack.
Use a paper clip to push the button inside the small
hole.
If the assembly isn’t done correctly, you may receive
one of the following error and warning messages. To
correct, follow the instructions as indicated within the
message.
Figure 6. Attaching the sensor module to the EM Eye
Meter
Figure 7. Location of the reset button at the bottom of
the EM Eye Meter