Lipid Profile TC TG
A codechip is included in each test strip package. Store the lipid test strips together with the codechip in the package.
Glucose Test Strip
A codechip is not necessary for glucose testing.
2. Preparing for measurement
Lipid Test System - components required:
•
STANDARD LipidoCare Analyzer
•
Lipid test strips for the desired parameter with the related codechip
•
Lancets and Lancing device
•
Tool for collecting the blood such as the STANDARD Ezi tube+ or micropipette
Only use STANDARD LipidoCare Lipid Test Strip and either STANDARD LipidoCare Blood Glucose Test Strip or SD Codefree Blood Glucose test
strip.
Precaution
Do
•
Read the test strip package insert.
•
Operate the Analyzer at the acceptable test-specific temperatures.
-
For Lipid testing 18-32 °C (64-90°F)
-
For Glucose testing 10-45 °C (50-113°F)
•
Place the Analyzer on a level surface or hold it steady in your hand.
•
Keep the guidelines for cleaning.
•
Make sure that all display elements are displayed.
•
Keep the STANDARD Lipid Check Strip away from sunlight.
Do not
•
Use lipid test strips beyond the expiration date, this may cause inaccurate results.
•
Touch or remove the test strip during actual measurement.
•
Subject the Analyzer to sudden movements during a measurement.