SMT Screen
The state machine trigger screen allows synchronization of the
modulation events with an externally provided hardware trigger (see Synthesizer
Module Font Panel), setting the CS-1 to operate as a state machine. The
operation as a state machine includes up to 10 states, each one incremented on
every rising edge of the hardware trigger. When the state number 10 is reached
the next hardware trigger resets the machine to state 1 and the cycle can
continue indefinitely. The number of states or events in a sequence is
programmed to be 1 through 10. Each state is identified by four parameters:
frequency, phase and amplitude of the RF carrier and the state of the RF relay
(on or off).
Note
:
it is not recommended to switch the RF relay with every trigger event because will
significantly limit the RF relay lifetime, which is specified by a finite number of switching
cycles. MTBF is typically 5 million cycles.
DISPLAYS
State n/m
Indicates which state is being edited (n) and the total number of
states that have been programmed (m).
RF ON/OFF
Current position of the RF relay at the 9.1926xx GHz output port.
Frequency
Current frequency of the 9.1926xx GHz output carrier expressed in
Hz.
Phase
Current phase of the 9.1926xx GHz output carrier expressed in
degrees.
Amplitude
Current amplitude of the 9.1926xx GHz output carrier expressed in
dBm.
MENU
ON
Enables, disables and resets the operation of CS-1 as a state machine
OFF
Disables the state machine operation.
RST
Resets the state machine cycle, forcing the next executable
state to be state 1.
SDI