Manual revision 041
Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
36
Status
Commands (
continued
)
DC
Display calibration coefficients, which are
initially factory-set and should agree with
Calibration Certificates shipped with
MicroCAT.
Example:
MicroCAT with a pressure sensor (user input in bold,
command used to modify parameter in parentheses
).
DC
SBE37SM-RS232 V 5.0.0 30051
temperature: 04-aug-17
[
TCalDate=
]
TA0 = 6.947802e-05
[
TA0=
]
TA1 = 2.615233e-04
[
TA1=
]
TA2 = -1.265233e-06
[
TA2=
]
TA3 = 1.310479e-07
[
TA3=
]
conductivity: 04-aug-17
[
CCalDate=
]
G = -1.00
[
CG=
]
H = 1.444342e-01
[
CH=
]
I = -3.112137e-04
[
CI=
]
J = 3.005941e-05
[
CJ=
]
CPCOR = -9.570001e-08
[
CPCor=
]
CTCOR = 3.250000e-06
[
CTCor=
]
WBOTC = 1.968100e-05
[
CWBOTC=
]
pressure S/N 2478619, range = 2901 psia, 03-aug-17
[
PRange=
(psi),
PCalDate=
]
PA0 = 0.00
[
PA0=
]
PA1 = 0.00
[
PA1=
]
PA2 = 0.00
[
PA2=
]
PTCA0 = 0.00
[
PTCA0=
]
PTCA1 = 0.00
[
PTCA1=
]
PTCA2 = 0.00
[
PTCA2=
]
PTCB0 = 0.00
[
PTCB0=
]
PTCB1 = 0.00
[
PTCB1=
]
PTCB2 = 0.00
[
PTCB2=
]
PTEMPA0 = 0.00
[
PTempA0=
]
PTEMPA1 = 0.00
[
PTempA1=
]
PTEMPA2 = 0.00
[
PTempA2=
]
POFFSET = 0.00
[
POffset=
(decibars)]
Notes:
The
DC
and
GetCC
responses
contain the same information, but in
different formats.
Dates shown are when calibrations
were performed.