Manual revision 041
Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
55
.
10.
After the data has been uploaded, Seaterm232 prompts you to run
SBE Data Processing’s Data Conversion module if desired. Data
Conversion converts the .hex (raw data) file to a .cnv file, which can then
be processed by other modules in SBE Data Processing.
A.
If you click Yes, Seaterm232 opens SBE Data Processing’s Data
Conversion module, and fills in the appropriate instrument
configuration (.xmlcon) file and data (.hex) file on the File Setup tab.
Notes
:
Ensure all data has been uploaded
from the MicroCAT by reviewing the
data in SBE Data Processing.
If you do not run Data Conversion
now, you can run it later by opening
SBE Data Processing.
See the SBE Data Processing
manual and/or Help for details.
Location to store all setup
information. Default is directory
with SeatermV2 application data,
when Data Conversion is
launched from Seaterm232.
Instrument configuration (.xmlcon)
file location, which is created by
Seaterm232, and contains
MicroCAT’s calibration coefficients
(see dialog box below).
Directory and file name for raw
data (.hex) file created by
Seaterm232 from uploaded data.