Manual revision 041
Section 4: Deploying and Operating MicroCAT
SBE 37-SM RS-232
58
11.
Once the data is converted to a .cnv file, use the other SBE Data
Processing modules as desired:
Derive module - Calculate additional derived variables.
Sea Plot module - Plot data.
Notes
:
To prepare for re-deployment:
1. After all data has been uploaded,
send
InitLogging
. If this is not sent,
new data will be stored after the last
sample, preventing use of the entire
memory.
2. Do
one
of the following:
Send
QS
to put the MicroCAT in
quiescent (sleep) state until ready
to redeploy. Quiescent current is
only 30 microAmps, so the battery
pack can be left in place without
significant loss of capacity.
Use
StartNow
to begin logging
immediately.
Set a date and time for logging to
start using
StartDateTime=
and
StartLater
.