Manual revision 041
Index
SBE 37-SM RS-232
84
Continued from previous page
027
04/08
Update for Version 3 firmware changes: many commands changed, power specifications changed.
Change stability specification for pressure to per year instead of per month.
Update connector maintenance information for consistency with application note 57.
Add information that
POffset
is in decibars.
028
07/08
Firmware revision 3.0c:
- If no P sensor is installed, reference pressure now appears in
GetCD
instead of
GetCC
, and
appears in
DS
instead of
DC
.
- Minimum conductivity frequency line removed from
GetCD
response if
PumpInstalled=n
.
-
SetPCBSerialNum1=
, etc. removed.
Update battery installation procedure, specifications, endurance, and shipping instructions for new
battery packs (12 AA lithium cells).
Correct cable length and external power example.
029
08/08
Firmware revision 3.0f:
- Add new output format to match firmware < 3.0.
DDb,e
now uploads in new format.
- If
StartLater
>90 days in future, does
StartNow
.
Manufacturing change: for plastic housing, 2 phillips-head screws at connector end cap end and 1
at sensor end cap end are replaced with hex screws. 9/64” allen wrench shipped with instrument.
030
01/09
Update for SeatermV2 terminal program.
Add information about compatibility with Vista.
Correction: Add PTempA0, PTempA1, PTempA2 to calibration coefficient commands.
031
01/10
Add information about Deployment Endurance Calculator.
Change Seasoft-Win32 to Seasoft V2, update file name to SeasoftV2_
date
.exe.
SBE Data Processing 7.20a: Add information about .xmlcon file.
Add CE mark.
Update SBE address.
Update anti-foul label in Appendix with new Container Handling requirement and new address.
032
07/10
Firmware 3.0j: Fix bug related to
StartDateTime=
. Previously, when
StartLater
was sent,
register ignored month in
StartDateTime
, and started at the next day and time corresponding to
the day and time (example: if it is July 1 and you set it to start on August 15, it ignored August
part of date, and started July 15). Documentation said could be started 90 days out, but this
actually limited it to 30 days out. Now, it provides a message saying it will start logging in
5 seconds if start date is more than 30 days out.
Add 60053 spares kit for plastic housing.
033
10/10
Update for changes to SeatermV2 version 1.1 (upload now converts .xml file to .hex and .xmlcon
files, which are used in Data Conversion to convert to .cnv file for further processing).
Remove references to Druck sensors (pressure sensors can be supplied by other manufacturers).
034
03/11
SeatermV2 1.1b changes:
- Update upload procedure, Seaterm232 automatically starts SBE Data Processing after upload
- Update SeatermV2 Instruments list screen capture
Add information about compatibility with Windows 7
035
08/12
Update Shipping Precautions for latest IATA rules.
Add Declaration of Conformity.
Add cable and wiring diagrams.
Add more information on selecting output variables for data processing of uploaded data.
Triton – update company name (Avantor Performance Materials) and link.
Remove factory-set commands:
SetPressureInstalled=
,
SetMfgDate=
,
SetPCBAssembly=
,
SetPCBSerialNum=
.
Glossary - Add information on ODO MicroCATs
Sampling Mode examples -- Add setting date/time to all examples, and add setting output format
to sync mode example.
Update configuration dialog box in SBE Data Processing.
Clarify information on correct battery pack to use with this MicroCAT.
Fix typos.
Continued on next page