background image

ADB06

Issue no: 1 - 19/06/13

6.4  

In-field Calibration Procedure

 

6.4.1 

ADB02 In Stand by Mode

6.4.2

  Press the CAL Key for 2 Seconds

 

ADB02 enters calibration mode, shows 

 

”cAl” and displays the calibration 

 

factor in use instead of total. the 

 

words “FAct” and “UsER” indicate 

 

which of the two factors is currently in use.

6.4.3 

  Press the RESET Key for 2 Seconds

 

ADB02 shows “cAl” and the partial at 

 

zero. ADB02 is ready to perform

 

on-site calibration.

6.4.4 

Dispensing into Sample Container

 

Without pressing any KEY, 

 

start dispensing into the 

 

sample container.

 

Dispensing can be interrupted and started again at will.  

 

continue dispensing until the level of the fluid in the sample  

 

container has reached the graduated area. there is no need  

 

to reach a preset quantity.

 

               

Indicated value 

 Real value

6.4.5

  Press the RESET Key

 

ADB02 is informed that the calibration 

 

dispensing operation is finished.

 

Make sure dispensing is correctly 

 

finished before performing this

 operation.

 

to calibrate the ADB02, the value indicated by the partial  

 

totaliser (example 9.800) must be changed to the real value  

 

marked on the graduated sample container.

 

In the bottom left part of the display an arrow appears  

 

(upwards and downwards), that shows the direction   

 

(increase or decrease) of the UsER K FActoR value change  

 

when the operations 5.4.6 or 5.4.7 are performed.

6.4.6

  Press the RESET Key

 

Arrow direction changes. the operation 

 

can be repeated if necessary.

6.4.7

  Press the CAL Key

 

the indicated value changes in the 

 

direction indicated by the arrow by one 

 

unit for each press of the cAl key. 

 

By continuous holding of the cAl key 

 

the first 5 units will change slowly and 

 

will then change quickly.

 

If the desired value is exceeded, repeat the operations from  

 

point (5.4.6).

6.4.8

  Press the RESET Key for 2 Seconds

 

ADB02 is informed that the calibration 

 

procedure is finished. Before doing this, 

 

make sure the DISPLAYED factor is the

 

ActUAl factor.

 

               

Indicated value  

Real value

 

ADB02 calculates the new UsER K FActoR. this calculation  

 

could require a few seconds, depending on the correction to  

 

be made. During this operation the arrow disappears but the  

 

cAl indication remains.

 

If this operation is performed after operation (5.4.5), without  

 

changing the indicated value, the UsER K FActoR would be  

 

the same as the FActoRY K FActoR, thus it is ignored.

6.4.9

 

At the end of the calculation, the new 

 

UsER K FActoR is shown for a few 

 

seconds, after which the restart cycle 

 

is repeated to finally achieve standby 

 condition.

ATTENTION! 

From now on, the indicated factor will become  

 

the calibration factor used by the meter and will continue to  

 

remain such even after a battery change.

6.4.10

 

ADB02 stores the new calibration factor 

 

and is ready for dispensing, applying the 

 

newly defined UsER K FActoR.

6.5  

Direct Modification of K Factor

 

this procedure is especially useful to correct a “mean error”  

 

obtainable on the basis of several performed dispensing  

 

operations. If normal ADB02 operation shows a mean

 

percentage error, this can be corrected by applying to the  

 

currently used calibration factor a correction of the same  

 

percentage. In this case, the percentage correction of the  

 

UsER K FActoR must be calculated by the operator in the  

 

following way:

 

Example:

 

Error percentage found E% - 0.9 %

 

CURRENT calibration factor 1,000

 

New USER K FACTOR = 1,000 * [(100 – ( - 0.9))/100]=

 

1,000 * [(100 + 0.9)/100] = 1.009

 

If the meter indicates less than the real dispensed value  

 

(negative error) the new calibration factor must be higher than  

 

the old one as shown in the example. the opposite applies if  

 

the meter shows more than the real dispensed value (positive

 error).

6.5.1

  ADB02 in STAND BY Mode

6.5.2

  Press the CAL Key for 2 Seconds

 

ADB02 enters calibration mode, shows 

 

“cAl” and displays the calibration factor 

 

being used instead of the partial. the 

 

words “FAct” and “UsER” indicate which 

 

of the two factors (factory or user) is 

 

currently being used.

6.5.3

  Press the RESET Key for 2 Seconds

 

 

ADB02 shows “cAl” and the partial at 

 zero.

 

ADB02 is ready to perform on-site 

 

calibration by dispensing.

    Original Language Version 

© Jack sealey limited

Summary of Contents for Adblue ADB06

Page 1: ...ORRECTLY AND WITH CARE FOR THE PURPOSE FOR WHICH IT IS INTENDED FAILURE TO DO SO MAY CAUSE DAMAGE OR PERSONAL INJURY AND WILL INVALIDATE THE WARRANTY PLEASE RETAIN THESE INSTRUCTIONS FOR FUTURE USE Thank you for purchasing a Sealey product Manufactured to a high standard this product will give you years of trouble free performance if these instructions are carefully followed and the product is cor...

Page 2: ...3 Indication of calibration mode 4 Totals register 6 figures with moving comma FROM 0 1 to 999999 that can indicate two types of Total 4 1 General Total that cannot be reset TOTAL 4 2 Resettable total Reset TOTAL 5 Indication of total multiplication factor x10 x100 6 Indication of type of total TOTAL Reset TOTAL 7 Indication of unit of measurement of Totals L Litres Gal Gallons 8 Indication of Flo...

Page 3: ...nd hold down the CAL button Why enter the calibration phases To display the currently used calibration factor Return to factory k factor after a previous calibration with user k factor Change the calibration factor using one of the two previously indicated procedures In calibration mode the partial and total dispensed quantities indicated on the display screen take on different meanings according ...

Page 4: ...Real value ADB02 calculates the new USER K FACTOR This calculation could require a few seconds depending on the correction to be made During this operation the arrow disappears but the CAL indication remains If this operation is performed after operation 5 4 5 without changing the indicated value the USER K FACTOR would be the same as the FACTORY K FACTOR thus it is ignored 6 4 9 At the end of the...

Page 5: ...ogether with the unit of measurement set at that time in this example Litres Litres 6 6 3 Press the RESET key to select the desired combination of unit of measurement amongst those shown below 6 6 4 Save the new combination by pressing the CAL key at length ADB02 will pass through the start cycle and will then be ready to dispense in the set units q WARNING The Resettable Total and Total registers...

Page 6: ...gue on 01284 757525 and leave your full name and address including your postcode Under the Waste Batteries and Accumulators Regulations 2009 Jack Sealey Ltd are required to inform potential purchasers of products containing batteries as defined within these regulations that they are registered with Valpak s registered compliance scheme Jack Sealey Ltd s Batteries Producer Registration Number BPRN ...

Reviews: