background image

5.  MAINTENANCE

4.  SPECIFICATION

   

WARNING!

 Before attempting to open the case, ensure that the test leads have been disconnected from the multimeter and that it is   

 

switched off to avoid electric shock hazard.

5.1. 

Changing a fuse.The fuses are located on the back of the circuit board. To gain access to the fuses, remove the protective rubber boot and  

 

the two screws from the rear of the multimeter. Lift off the rear cover, replace the appropriate fuse and re-assemble in reverse order.

   

WARNING!

 

ALWAYS

 replace a fuse with one of the correct rating.

 

Fuse 1: 500mA/250V

 

Fuse 2: 10A/250V

5.2. 

If the battery sign         appears on the LCD display, it indicates that the battery must be replaced. repeat the steps detailed in section  

 

5.1 to remove the rear cover, replace the battery (9V PP3) and re-assemble in reverse order. 

5.3.

 

Clean the multimeter's casing using a slightly dampened cloth and mild detergent - do not use any abrasives or solvents. Clean the inside  

 

of each terminal using a swab soaked in isopropyl alcohol, use a new swab to apply a light coat of machine oil to each terminal.

5.4.

 

If the multimeter is to be stored for a long period of time, remove the battery first to avoid any damage. 

DC VOLTAGE

Range

Accuracy

Resolution

200mV

±0.8% of reading ±4 digits

100

µV

2V

±0.8% of reading ±4 digits

1mV

20V

±0.8% of reading ±4 digits

10mV

200V

±0.8% of reading ±4 digits

100mV

600V

±1.2% of reading ±4 digits

1V

Input impedance: 10M

 for all ranges. overload 

Protection: 600V dc or peak ac on all ranges.
overload Protection:200V rms ac for 200mV range and 
600V dc or ac for other ranges.

AC VOLTAGE

Range

Accuracy

Resolution

2V

±0.8% of reading ±5 digits

1mV

20V

±0.8% of reading ±5 digits

10mV

200V

±0.8% of reading ±5 digits

100mV

600V

±1.2% of reading ±4 digits

1V

Input impedance: 10M

 for all ranges.

Frequency range: 40Hz to 1kHz
overload Protection: 220V rms ac for 200mV range and 
600V dc or ac for other ranges.
Indication: Average (rms of sine wave).

DC CURRENT

Range

Accuracy

Resolution

2mA

±0.8% of reading ±4 digits

1µA

20mA

±0.8% of reading ±4 digits

10µA

200mA

±1.2% of reading ±5 digits

100µA

10A

±2.0% of reading ±5 digits

10mA

overload protection: Fuse 1: 500mA/250V  
Maximum Input Current: Fuse 2: 10A/250V (10A 
up to 15 seconds).
Measuring Voltage Drop: 200mV.   

AUDIBLE CONTINUITY TEST

Range

Description

Buzzer sounds if 

resistance is less than 30 ohms

open Circuit Voltage approximately 2.8V 

BATTERY TEST

Range

Accuracy

Current

1.5V

±0.8% of reading ±2 digits

60mA

9.0V

±0.8% of reading ±2 digits

12mA

TRANSISTOR hFE TEST (External test socket)

Range

Description

Test Condition

hFE

Displays approximate 
hFE value (1-1000) of 
transistor.

Base current 

approximately 10µA.

VCE approximately 

2.8V.

AC CURRENT

Range

Accuracy

Resolution

2mA

±1.2% of reading ±3 digits

1µA

20mA

±2.0% of reading ±3 digits

10µA

200mA

±2.0% of reading ±3 digits

100µA

10A

±3.0% of reading ±7 digits

10mA

overload protection: Fuse 1: 500mA/250V  
Indication: Average (rms of sine wave).

 

Maximum Input Current: Fuse 2: 10A/250V (10A 
up to 15 seconds).
Measuring Voltage Drop: 200mV.
Frequency range: 40Hz to 1kHz

.

RESISTANCE

Range

Accuracy

Resolution

200

±1.0% of reading ±8 digits

0.1

2k

±1.2% of reading ±8 digits

1

20k

±1.2% of reading ±8 digits

10

200k

±1.2% of reading ±8 digits

100

2M

±1.2% of reading ±8 digits

1k

20M

±1.2% of reading ±8 digits

10k

Maximum open circuit voltage less than 750mV.
overload protection: 220V dc/ac rms on all ranges.

 

  Environmental Protection.

 

  

recycle unwanted materials instead of disposing of them as waste. All tools, accessories and packaging should be sorted, 

 

  

taken to a recycle centre and disposed of in a manner which is compatible with the environment.

 
 

  

When the product is no longer required, it must be disposed of in an environmentally protective way. 

Battery Removal.

See Section 5.2.

1. 

remove and dispose of according to local authority guidelines.

2. 

 

NOTE: It is our policy to continually improve products and as such we reserve the right to alter data, specifications and component parts without prior notice.
IMPORTANT: 

no liability is accepted for incorrect use of this product. 

WARRANTY: 

Guarantee is 12 months from purchase date, proof of which will be required for any claim. 

INFORMATION:

 For a copy of our latest catalogue and promotions call us on 01284 757525 and leave your full name and address, including postcode.

Sole UK Distributor, Sealey Group, 

Kempson Way, Suffolk Business Park

Bury St. Edmunds, Suffolk,
IP32 7Ar

Original Language Version

01284 757500

01284 703534

sales@sealey.co.uk

www.sealey.co.uk

Web

email

  MM24        Issue no: 2 - 02/07/10

DIODE TEST 

Range

Description

Test Condition

  (2k)

Displays approximate 

forward Voltage of 

Diode

Forward DC current 
approximately 1mA.

reversed DC voltage 

approximately 2.8V