PANEL METER FOR INW PT12 SENSOR
Seametrics • 253.872.0284
Page 9
inwusa.com
Register Number
The meter reads a particular register (memory) address on the sensor in order to display
a reading. The PT12 has three registers - one for pressure, one for temperature, and
one for voltage. The meter comes preconfigured to read pressure. If you want to read
temperature, you must set the meter to read the correct register, as shown below. (Note,
if you are using other than PSI as display units for pressure, see the following section
on display units.)
Display Units
The panel meter comes preconfigured to display readings in the units you have
ordered for your PT12. The PT12 default is PSI for pressure and degrees Celsius for
temperature, unless something else is specified in your order.
If you need to change units, see Appendix A.
NOTE: With firmware version
0.13 and higher you can also
read from registers:
462593 for pressure
462595 for temperature
462597 for voltage