WMP-SERIES INSTRUCTIONS

Seametrics • 253.872.0284 

Page 5 

seametrics.com

Step 1

Position gasket 

at either end

Step 2

Place adapter 

against gasket, 

open screw clamps 

to clear flanges

Step 3

Place screw-clamps 

over both adapter 

and meter flanges 

and tighten screws

Positioning

This is an all position meter which can be installed either vertically or horizontally, register up, down, or angled. However, 

entrained air or solids may make some positions preferable to others. 

Recommended:

May prevent sediment

or bubble problems

Electrode moved 

from top by 

rotating meter

Intermittent air

bubbles

miss electrode

Electrodes free 

from sediment 

build-up

WMP

NOTE:  Above installation instructions are for WMP101/104-200 (2”) and WMP101/104-300 (3”) only.  

The WMP101/104-100 (1”) is provided with integrated NPT female threaded ends.

INSTALLATION

Piping Conditions

It is highly recommended that the meter be installed with 

a length of straight pipe at least two times the diameter 

upstream  and  one  diameters  downstream.  Some  piping 

conditions require more than this. See next page for 

recommendations.

End Connections

The  meter  comes  with  Banjo™  union-type  flange 

connections for ease in servicing the meter. To connect 

these to piping ends, a variety of kits are available from any 

Banjo dealer or from Seametrics.

Follow the diagrams below to make the connections.

Not ideal:

Air bubbles and sediment on the

electrodes can affect accuracy

Intermittent air 

bubbles pass over 

electrode

Possible 

sediment 

build-up

WMP

Summary of Contents for WMP Series

Page 1: ...PROUDLY MADE IN THE USA C e rtified Compa n y ISO 9001 2008 WMP Series Plastic Bodied Magmeter Instructions 2 inch 1 inch 3 inch...

Page 2: ...ng is made of rugged powder coated diecast aluminum It can be fitted with cross drilled screws and seal wire for tamper evidence Flow rate and total can be displayed in a variety of units customer sel...

Page 3: ...Million Gallons Day Cubic Feet Second Cubic Feet Minute Liters Second Liters Minute Million Liters Day Cubic Meters Minute Cubic Meters Hour Acre Feet Acre Inch Gallons Gallons x 1000 Cubic Feet Lite...

Page 4: ...5 50 4 625 9 50 3 25 2 00 5 50 2 03 6 03 4 625 10 75 4 40 3 00 7 30 7 30 2 625 3 Inch Dimensions Dimensions shown without the Seametrics WMP Fitting Kit Flow Range 1 2 3 Gal Min Liter Sec Gal Min Lite...

Page 5: ...P NOTE Above installation instructions are for WMP101 104 200 2 and WMP101 104 300 3 only The WMP101 104 100 1 is provided with integrated NPT female threaded ends INSTALLATION Piping Conditions It is...

Page 6: ...lem Allows air pockets to form at sensor Better Installation Ensures full pipe Possible Problem Post valve cavitation can create air pocket Better Installation Keeps pipe full at sensor Air can be tra...

Page 7: ...umulated total in the units for which it was ordered The top line is the total the bottom line is the rate and indicators give the units ac ft GPM for instance Empty or partially full pipe is automati...

Page 8: ...efully unplug each of the two battery wiring harness connectors Slip the batteries out from under the elastic retainer and replace them with the new batteries Reconnect each of the two wiring harness...

Reviews: