PrimeTest 200
Operating Instructions
6
7.4
Voltage measuring range
Measuring Range
Resolution
Accuracy
Overload Protection
250 V AC/DC
1V
±2% +2 digits
300 V AC/DC
Frequency Range
up to 400Hz
Analogue bargraph display
0 to 250 V
Bargraph accuracy
10%
Input protection
>20M
Ω
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the voltage measurement function. A repetitive audible alarm
is activated and a flashing high voltage symbol is displayed on the LCD.
7.5
IEC lead test
The IEC lead test is a sequence of earth continuity measurement (section 7.1)
- insulation resistance measurement (section 7.2) - Live/Neutral checks for o/c,
s/c and crossed conductors in the IEC lead.
Test voltage
9 VDC
Test duration
10 seconds
7.6
Temperature measuring range °C
Measuring Range
Resolution
Accuracy
Overload Protection
-50°C to +400°C
0.1°C
±5% +2digits
300 V AC/DC
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the temperature measurement function. A repetitive audible
alarm is activated and a flashing high voltage symbol is displayed on the LCD.
7.7
Touch current measurement
Measuring Range
Resolution
Accuracy
Overload Protection
3.5mA
0.01mA
±10% + 2 digits (5° - 40°)
N/A
±15% + 2 digits (0° - 5°)
Test voltage
mains supply @ 230V - 15%/+10% @ 10A max
Frequency
50Hz ±1%
Test duration
2 seconds minimum
Frequency response
IEC 61010-1 Annex A.1
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the touch current test function. A repetitive audible alarm is
activated and a flashing high voltage symbol is displayed on the LCD.
The test is also disabled when phase-neutral (PN) or phase-earth (PE) voltage
is not at mains potential or a voltage of 40 V AC or greater exists between
neutral-earth (NE).
7.8
Differential current measurement
Measuring Range
Resolution
Accuracy
Overload Protection
19.99m A
0.01mA
±5% + 2 digits
N/A
Test voltage
mains supply @ 230V - 15%/+10% @ 10A max
Frequency
50Hz ±1%
Test duration
2 seconds minimum
Frequency response
IEC 61010-1 Annex A.1 (from 40Hz)
An input voltage of
≥
30 V AC/DC, detected between the + and - input measuring
terminals, disables the differential current test function. A repetitive audible
alarm is activated and a flashing high voltage symbol is displayed on the LCD.
The test is also disabled when phase-neutral (PN) or phase-earth (PE) voltage
is not at mains potential or a voltage of 40 V AC or greater exists between
neutral-earth (NE).
Summary of Contents for PrimeTest 200
Page 3: ...Operating Instructions PrimeTest 200 Fig 5 Fig 6...
Page 4: ...Operating Instructions PrimeTest 200 Fig 7 Fig 8...
Page 5: ...Operating Instructions PrimeTest 200 Fig 9 Fig 10...
Page 6: ...Operating Instructions PrimeTest 200 Fig 11...
Page 7: ...Operating Instructions PrimeTest 200 Fig 12B Fig 12A...
Page 8: ...PrimeTest 200 Operating Instructions...
Page 10: ...PrimeTest 200 Operating Instructions...