14
8.9
Temperature Measurement Function, °C.
See Figure 11
Attention - Disconnect the IEC mains cable from the IEC socket
before commencing this test. Electric Shock Danger!
Do not exceed the maximum permitted voltage of 30 V AC/DC with
respect to earth potential! Electric shock danger!
8.9.1
Select the temperature measurement function on the rotary switch 10 .
The LCD will display two probe symbols to indicate that both the red
and black sockets require connections. Ensure the IEC mains lead is
disconnected from the PrimeTest 200.
8.9.2
Plug the PrimeTest 200 temperature module (see 3.2.1) into the red 13
and black 12 sockets on the PrimeTest 200.
8.9.3
Plug the type K thermocouple into the PrimeTest 200 temperature module.
Select the switch position to °C on the top of the temperature module.
8.9.4
The large digital display 2 indicates the temperature measurement in
°C from -50 to 400°C.
Where the measured temperature value is greater than +400°C the
large display indicates >400°C. Where the measured temperature
value is less than -50°C the large display indicates <-50°C
8.10
Voltage Measuring Function,
See Figure 12a and 12b
8.10.1 Select the voltage measurement function on the rotary switch 10 .
8.10.2 Plug the voltage measurement adaptor into the IEC mains input plug 14
on the end of the PrimeTest 200.
8.10.3 Plug the black test lead into the negative, black socket on the voltage
measurement adaptor.
8.10.4 Plug the red test lead into the positive, red socket on the voltage
measurement adaptor.
8.10.5 Connect the red and black test leads to the point to be measured.
8.10.6 The large display 2 indicates the digital and analogue voltage
measurement in V AC/DC up to 300 V AC/DC.
Where the measured voltage value, is greater than 250 V AC/DC the
large display indicates > 250 V AC/DC for over range.
8.10.7 An input voltage of
≥
30 V AC/DC, detected between the + and - input
measuring terminals, disables the voltage measurement function. A
repetitive audible alarm is activated and a flashing high voltage symbol
is displayed on the LCD.
The PrimeTest 200 indicates either a DC (direct) voltage or an AC
(alternating) voltage. If the measured voltage has a DC component
and an AC component, only the component which has the greater
magnitude will be measured and displayed.
9.0
Maintenance
Before opening the PrimeTest 200 ensure that it is disconnected
from all voltage! Electric shock danger!
PrimeTest 200
Operating Instructions
Summary of Contents for PrimeTest 200
Page 3: ...Operating Instructions PrimeTest 200 Fig 5 Fig 6...
Page 4: ...Operating Instructions PrimeTest 200 Fig 7 Fig 8...
Page 5: ...Operating Instructions PrimeTest 200 Fig 9 Fig 10...
Page 6: ...Operating Instructions PrimeTest 200 Fig 11...
Page 7: ...Operating Instructions PrimeTest 200 Fig 12B Fig 12A...
Page 8: ...PrimeTest 200 Operating Instructions...
Page 10: ...PrimeTest 200 Operating Instructions...