background image

Operating instructions HLE 500/5000

 

 

19 

MAN_HLE500_HLE5000_eng_02 

3.10  Location of pipes using the RSP-3 or PWG  

(HLE 5000 only) 

To  switch  the  HLE  5000  into  this  mode,  after  switching  on, 
soft key 2 (underneath the impulse symbol           ) should be 
pressed whilst the welcome image is displayed. 
The  HLE  5000  is  now  in  pipe  location  mode,  shown  by  the 
impulse symbol at the top edge of the display. 
This  mode  of  operation  is  particularly  suited  to  display 
impulse noise, such as the tapping of the pipe pecker or the 
PWG.  The  horizontal  level  bars  respond  particularly 
sensitively  to  impulses,  the  bar  is  shown  much  larger  and  is 
delayed  more  and  the  filters  are  set  to  lower  frequencies  Of 
course,  if  required,  the  filter  frequencies  can  be  altered  as 
usual.  
The HLE5000 stays in pipe location mode until it is switched 
off. After being turned on again, it will be in normal mode. 
 

3.11  Switching off 

 

The equipment is switched off by a longer press on the 

 

button.  Any  measurements  stored  will  be  lost.  After  35 
minutes 

in 

operation, 

the 

equipment 

switches 

off 

automatically.  
 
In  histogram  mode 

  the  equipment  will  also  switch  off 

automatically  after  35  minutes,  as  long  as  the  mute  and  the 
illumination button are not used.  
 

Summary of Contents for Hydrolux HLE 500

Page 1: ...Prospekten ohne Rahmen Ansprechpartner Frau Badziura Leitungsortung Line Locating Rohrleitungsnetze Water Networks Kommunikationsnetze Communication Networks Elektrizit tsnetze Power Networks Edition...

Page 2: ...Operating instructions HLE 500 5000 II MAN_HLE500_HLE5000_eng_02...

Page 3: ...uality standard DIN ISO 9001 and the European standard EN 29001 This quality system was approved by the German Association for the Certification of Quality Systems with the numbers EN 19677 and DQS 19...

Page 4: ...Operating instructions HLE 500 5000 IV MAN_HLE500_HLE5000_eng_02...

Page 5: ...1 Headphones and microphone connection 11 3 2 Switching on 11 3 3 Volume and amplification 12 3 3 1 Setting the volume 12 3 3 2 Setting the gain 13 3 4 The amplification display 13 3 5 Filter setting...

Page 6: ...with magnet 9 Fig 6 PAM U with 3 point foot 10 Fig 7 Start image with version number 11 Fig 8 Opening menu after switch on 12 Fig 9 Setting the volume 12 Fig 10 Setting the gain 13 Fig 11 Current and...

Page 7: ...the ground microphone a new minimum display results A comparison of the measured values is thus always possible In the HISTOGRAM function each measurement location is stored one by one This equipment...

Page 8: ...sides of the equipment and can be quickly plugged in or removed 1 3 Technical data Analysis bandwidth 0 Hz 4000 Hz Filter cut off frequencies 0 70 Hz 106 Hz 160 Hz 240 Hz 360 Hz 540 Hz 800 Hz 1200 Hz...

Page 9: ...le to PAM W 1 VK65 3 point foot adapter PAM W 1D Carrying pole for PAM W 1 Universal PAM U microphone PAM U Magnetic adapter Sensor rod 3 point foot for PAM U PAM U D Extension rod VST T 1 Ground micr...

Page 10: ...Optional accessories Radio module for wireless transmission between microphone and amplifier PAM 868 Sensor rod extension VST T 1 Special headphones highly insulated from extraneous noise KM2 Sliding...

Page 11: ...controls the devices which can be connected and the power supply 2 1 HLE 5000 500 controls Fig 1 HLE 5000 500 control panel 2 2 Connections at the side The sockets for the microphone and the headphone...

Page 12: ...or headphones and microphones from SebaKMT Connecting other headphones or microphones can lead to equipment failure or damage to the HLE 5000 500 2 3 Battery chamber After opening the base flap fig 3...

Page 13: ...hange the batteries change all 10 batteries at one time 2 5 Illumination of the LCD display By briefly pushing the on off button when the equipment is turned on the illumination for the display will b...

Page 14: ...ly well shielded from the wind using a bell shaped guard The carrying pole can be removed with a quarter turn to the left This is particularly useful in reducing extraneous noise to a minimum when the...

Page 15: ...the working posture is more comfortable but due to the extension wind and surrounding sources of noise are more intrusive Take care When pulling the microphone out always pull the body of the microph...

Page 16: ...round microphone PAM W 1 but nevertheless very good results can be achieved with this variant Fig 6 PAM U with 3 point foot 2 7 Headphones The KR 2 headphones with ambient sound insulation are supplie...

Page 17: ...both the headphones and the microphone The reverse is true when switching off Always switch the equipment off first and then remove the microphone and headphones 3 2 Switching on By pushing the on of...

Page 18: ...ped with separate controls for volume and amplification Thus every user can make the best settings for their personal characteristics 3 3 1 Setting the volume Using the two buttons on the right see fi...

Page 19: ...g 3 4 should be selected Changing the gain affects the level bar and the volume in the headphones 3 4 The amplification display The bar display fig 11 shows both the current value of the sound picked...

Page 20: ...erneath there is a horizontal bar which identifies the selected filter range The lower and upper cut off is shown to the left and right of the bars respectively 1800 4000 0 70 Fig 12 Filter settings T...

Page 21: ...t off frequencies you can set the 9 frequencies specified above and can thus easily set the special filter range for every measurement After the filter range has been correctly set use the soft key bu...

Page 22: ...should first of all push the mute button X see fig 1 This causes the sound in the headphones to be interrupted and the current level will be frozen This means that your hearing is protected as the lou...

Page 23: ...value which is of particular significance when looking for burst pipes Fig 15 Dual segment analysis DSA With the first push of the mute button X the current measurement is stored and simultaneously d...

Page 24: ...ording the flow noise of a valve To do that the microphone is put onto the pipe and the long term measurement is started Then you close the valve for a certain time at least 2 minutes and then open it...

Page 25: ...he PWG The horizontal level bars respond particularly sensitively to impulses the bar is shown much larger and is delayed more and the filters are set to lower frequencies Of course if required the fi...

Page 26: ...nd check the polarity of the batteries See fig 3 If the polarity of all of the batteries is OK the state of every single battery must be checked 4 3 No sound can be heard With working equipment there...

Page 27: ......

Reviews: