Operating instructions HLE 500/5000
2
MAN_HLE500_HLE5000_eng_02
1.2
Construction
The leak location equipment is in a splash-proof housing
made of robust plastic. The few control buttons are so
arranged that they can also be operated wearing gloves. The
display is equipped with backlighting. That means that work is
also possible under poor lighting conditions, or at night.
The batteries required to supply the power are housed in the
base of the equipment behind a cover which is easy to open
to quickly exchange the batteries. The connections for the
microphone and headphones are on the two sides of the
equipment and can be quickly plugged in or removed.
1.3
Technical data
Analysis bandwidth:
0 Hz – 4000 Hz
Filter cut-off frequencies:
0 - 70 Hz, 106 Hz, 160 Hz,
240 Hz, 360 Hz, 540 Hz,
800 Hz, 1200 Hz, 1800 –
4000 Hz
Histogram recordings:
9 dual displays
Storage of noise levels:
3 – 10 – 30 min
Display 130 x 36 mm:
LCD display
LCD illumination:
available
Power supply 1:
Power supply 2 (option):
10 x AA batteries, (1.5 Volt)
12 x rechargeable batteries
(1.2 Volt)
Operating time:
> 35 h (battery),
> 15 h (rechargeable battery)
Storage:
9 measurements
Mute button:
available
Operating temperature:
-10 to +50 °C
Storage temperature:
-10 to +70 °C
Protection class when in
operation:
IP 54
Dimensions (L x W x D):
215 x 95 x 110 mm
Weight: HLE 5000
Weight: PAM W-1
Weight: PAM U
1200 g (with batteries)
3.5 kg (with carrying pole)
500 g