46
Start of the storing
If it is necessary to synchronize the storing of measured values on several transmitters, the starting time of the storing can
be set.
Select the starting time of the storing of measured values.
Promptly
: Storing will be started immediately.
On full 5 min.
: Storing will be started on the next full 5 minutes.
On full 10 min.
: Storing will be started on the next full 10 minutes.
On quarter hour
: Storing will be started on the next full 15 minutes.
On half hour
: Storing will be started on the next half hour.
On full hour
: Storing will be started on the next full hour.
Example:
current time: 9:06
setting:
On full 10 min.
Storing will be started at 9:10.
Ringbuffer
The setting of ringbuffer affects the storing of measured values as soon as the data memory is full:
• If the ringbuffer is activated, the available data logger will be halved. The oldest measured values will be overwritten.
Only the data logger memory that was free during the activation will be used by the ringbuffer. If more data logger mem-
ory is necessary, measured values in the data logger should previously be deleted.
• If the ringbuffer is deactivated, the storing of measured values will be stopped.
Select the behavior of the ringbuffer. Press ENTER.
Storage mode
Select the storage mode. Press ENTER.
If
sample
is selected, the displayed measured value will be used for storing and online
transmission of data.
If
average
is selected, the average of all values measured during a storage interval will be
used for storing and online transmission of data.
Storing of the totalizers
see section 11.3
Storing of the signal amplitude
If
on
is selected and the data logger is activated, the amplitude of the measured signal will
be stored together with the measured values. Press ENTER.
Storing of the sound speed
If
on
is selected and the data logger is activated, the sound speed of the medium will be
stored together with the measured values. Press ENTER.
Storing of the diagnostic values
If
on
is selected and the data logger is activated, the diagnostic values will be stored to-
gether with the measured values. Press ENTER.
Note!
Storage mode
=
average
The average of the physical quantity and other physical quantities assigned to the measuring chan-
nel will be calculated.
If the storage rate < 5 s (see section 12.1.2) is selected,
sample
will be used.
If no average could be calculated over the complete storage interval, the value will be marked as in-
valid. The ASCII file will contain "
???
" instead of invalid average values of the physical quantity.
Note!
The storage mode does not affect the outputs.
Start logger
↕
Promptly
Ringbuffer
off
>ON<
Storage mode
>SAMPLE< average
Store Amplitude
off
>ON<
Store c-Medium
off
>ON<
Store diagnostic
off
>ON<