49
12.2.4 Transmission Parameters
• the transmitter sends CRLF-terminated ASCII
• max. line length: 255 digits
RS232
• default: 9600 bits/s, 8 data bits, even parity, 2 stop bits, protocol RTS/CTS (hardware, handshake)
The transmission parameters of the RS232 interface can be changed:
Enter HotCode
232-0-
(see section 8.4).
Set the transmission parameters in the 4 scroll lists. Press ENTER.
•
baud
: baud rate
•
data
: number of data bits
•
par
: parity
•
st
: number of stop bits
RS485
• default: 9600 bits/s, 8 data bits, even parity, 1 stop bit
The transmission parameters of the RS485 interface can be changed in the program branch
Special Funct.\SYSTEM
settings\Network
. This display will only be indicated if the transmitter has an RS485 interface.
Select
Special Funct.\SYSTEM settings\Network
to change the settings of the
transmission parameters.
Press ENTER to confirm the address of the measuring instrument in the network.
Select
default
to display the default transmission parameters.
Select
setup
to change the transmission parameters. Press ENTER.
Set the transmission parameters in the 3 scroll lists. Press ENTER.
•
baud
: baud rate
•
parity
: parity
•
st
: number of stop bits
The default transmission parameters will be set if
default
is selected and the transmis-
sion parameters have not been changed.
12.2.5 Online Transmission of Data to a Terminal Program
• Start the terminal program.
• Enter the transmission parameters into the terminal program (see section 12.2.4). The transmission parameters of the
terminal program and of the transmitter have to be identical.
• Select the program branch
Output Options
. Press ENTER.
• Select the channel for which the online transmission of data is to be activated. Press ENTER until the menu item
Serial
Output
is displayed.
Select
yes
to activate the online transmission of data.
• Set the storage rate (see section 12.1.2).
• Start the measurement. The measuring point number will be requested (see section 12.1.4).
The measured values are transmitted during the measurement.
baud<data par st
9600 8bit EVEN 2
SYSTEM
settings↕
Network
Device address:
0
ADR
RS485 protocol
default >SETUP<
>baud< parity st
9600 EVEN
1
Serial Output
no
>YES<
SEND ONLINE-HEAD
20
mm