65
15.1.3 Function Test
The function of the installed output can now be tested. Connect a multimeter with the installed output.
Test of the analog output
The current output is tested in the display. Enter a test value. It has to be within the output
range. Press ENTER.
If the multimeter displays the entered value, the output functions correctly.
Press
yes
to repeat the test,
no
to return to the
SYSTEM settings
. Press ENTER.
Test of the binary outputs
Select
Opto-Relay OFF
in the scroll list
Output Test
to test the de-energized state of
the output. Press ENTER. Measure the resistance at the output. The value has to be high
ohmic.
Select
yes
. Press ENTER.
Select
Opto-Relay ON
in the scroll list
Output Test
to test the energized state of the
output. Press ENTER. Measure the resistance at the output. The value has to be low ohm-
ic.
Select
yes
to repeat the test,
no
to return to
SYSTEM settings
. Press ENTER.
15.2
Error Value Delay
The error value delay is the time interval after which the error value will be transmitted to the output in case no valid mea-
sured values are available. The error value delay can be entered in the program branch
Output Options
if this menu
item has previously been activated in the program branch
Special Funct.
. If the error value delay is not entered, the
damping factor will be used.
Select
Special Funct.\SYSTEM settings\Dialogs/Menus\Error-val. de-
lay.
Select
Damping
if the damping factor is to be used as the error value delay. Select
Edit
to
activate the menu item Error-val. delay in the program branch
Output Options
.
From now on, the error value delay can be entered in the program branch
Output Op-
tions
.
15.3
Activation of an Analog Output
Press ENTER until
Current Loop
is displayed. Select
yes
to activate the output.
Press ENTER.
Note!
An output can only be activated in the program branch
Output Options
if it has previously been in-
stalled.
I1:Output Test
4
mA
I1= 4.0 mA
Again? no >YES<
B1:Output Test
↕
Opto-Relay OFF
B1=OFF
AGAIN? no >YES<
B1:Output Test
↕
Opto-Relay ON
B1=ON
AGAIN? no >YES<
Error-val. delay
>DAMPING< edit
Error-val. delay
10
s
Current Loop
I1: no
>YES<