80
INIT-
resistant
input of the cut-off flow for positive measured values
range: 0...12.7 cm/s (0.127 m/s)
default: 2.5 cm/s (0.025 m/s)
This display will only be indicated if
Cut-off Flow = sign
and
Cut-off Flow = user
have been selected.
X
input of the cut-off flow for negative measured values
range: -12.7...0 cm/s
default: -2.5 cm/s
This display will only be indicated if
Cut-off Flow = sign
and
Cut-off Flow = user
have been selected.
X
input of the cut-off flow for the absolute value of the measured values
range: 0...12.7 cm/s
default: 2.5 cm/s
This display will only be indicated if
Cut-off Flow = absolut
and
Cut-off Flow = user
has been selected.
X
input of the max. signal amplification
range: 0...255
0: no limit of the signal amplification
This display will only be indicated if the SuperUser mode is activated.
X
input of the fixed upper limit of the sound speed
range: 0…3000 m/s
0: the default value 1 848 m/s is used
This display will only be indicated if the SuperUser mode is activated.
X
input of the offset
range: 0…900 m/s
0: the default value 300 m/s is used
This display will only be indicated if the SuperUser mode is activated.
X
activation of the overflow of the totalizers
X
activation of the taking-over of the totalizer values after a restart of the
measurement
X
input of the time interval without any valid measured values after which the
transmitter recognizes a long measurement failure
0: the default value 30 s is used
This display will only be indicated if the SuperUser mode is activated.
X
input of the number of decimal places for the totalizers
Automatic
: dynamic adjustment
Fixed to x digit
: 0...4 decimal places
This display will only be indicated if the SuperUser mode is activated.
X
activation of the manual reset of the totalizers
This display will only be indicated if the SuperUser mode is activated.
X
Bad soundspeed
thresh. 2007 m/s
Bad soundspeed
offset: +321 m/s
Quant. wrapping
off
>ON<
Quantity recall
off
>ON<
Gain threshold
Fail if > 90 dB
Do not total. if
no meas.> 0 s
Total digits
↕
Automatic
3xC clear totals
off
>ON<
+Cut-off Flow
2.5
cm/s
-Cut-off Flow
-2.5 cm/s
Cut-off Flow
2.5
cm/s