81
INIT-
resistant
activation of the display of the sum of the totalizers.
This display will only be indicated if the SuperUser mode is activated.
X
activation of the display of the last valid measured value
This display will only be indicated if the SuperUser mode is activated.
X
activation of the turbulence mode
X
Special Funct.\SYSTEM settings\Measuring\Calibration
definition of the profile bounds
factory
: the default profile bounds is used
user
: the profile bounds can be defined
This display will only be indicated if the SuperUser mode is activated.
input of the max. Reynolds number at which the flow is laminar
range: 0...25 500 (rounded to hundreds)
0: the default value 1 000 is used
This display will only be indicated if the SuperUser mode is activated
and
Pro- file bounds
=
user
is selected.
input of the min. Reynolds number at which the flow is turbulent
range: 0...25 500 (rounded to hundreds)
0: the default value 3 000 is used
This display will only be indicated if the SuperUser mode is activated and
Pro- file bounds = user
is selected.
request if an additional correction of the flow velocity is to be defined
on
: the correction data can be defined
off
: no correction of the flow velocity will be used
This display will only be indicated if the SuperUser mode is activated.
input of the slope for the correction formula
range: -2.000...+2.000
0: no correction
This display will only be indicated if the SuperUser mode is activated and
Cali- bration = on
is selected.
input of the offset
range: -12.7...+12.7 cm/s
0: no offset
This display will only be indicated if the SuperUser mode is activated and
Cali- bration = on
is selected.
SYSTEM settings\Proc. outputs
selection of the displays for the setting of the outputs of the transmitter
selection of the output to be installed
Keep display val
off
>ON<
Turbulence mode
off
>ON<
Profile bounds
factory >USER<
Laminar flow
if R*<
0
Show
Σ
Q
off
>ON<
Turbulent flow
if R*>
0
Calibration ?
>OFF<
on
Slope=
1.00
Offset=
0.0
cm/s
SYSTEM settings↕
Proc. outputs
Install Output ↕
Current I1