APR TECHNIQUE CHANGES
The APR techniques are factory pre-programmed to standard technique sets.
All parameters of the APR techniques may be manually modified by the
operator and stored in the non-volatile memory for later use.
If the operator determines that some factors in an APR technique should be
re-programmed, use the following procedure:
1.
Select an APR technique and modify the factors and selections of
Workstations or AEC which require to be re-programmed.
2.
Verify that all factors of the technique are at the required values.
3.
Simultaneously press the push-buttons 2--6 of the APR Display to store
the new technique. (
Refer to Illustration 3-2
.)
The newly selected technique is now stored in memory and can be
recalled for future examinations.
This procedure only changes the technique values of the selected
patient size, repeat the procedure for other patient sizes.
Illustration 3-2
APR Storage Push-buttons