30
M731400M00
CMRR (Common Mode Rejection Ratio) :
The CMRR is the ability of a multimeter to reject the Common
Mode Voltage
Vcm
(The voltage present on both the
COM
and
VOLTAGE
input terminal
with respect to earth ground). The
Vcm normally comes from the electromagnetic interference of
high voltage power source line or generators.
NMRR (Normal Mode Rejection Ratio) :
The NMRR is the ability of a multimeter to reject the
Unwanted AC noise, V
NM
, in DC measurement.
Burden Voltage :
Burden Voltage (V
BURDEN
) is voltage present on the
CURRENT
input terminal
and
COM
terminal
of a multimeter. The presence
of the Burden Voltage on the current under test flows through
the impedance of the current sensing circuit of the multimeter .
The Burden Voltage will cause the measuring value to be less
than the actual value. For accurate measurements use the
approxi-mation described in the operation of Measuring
Current.