Page 5.1
5.
TRIGGERS
This chapter describes all possible triggers of the appliance.
They are used by:
-
the « setup » menu, with the A and B alarms
-
the « triggers » menu, with the start and stop parameters of the recording.
Edge/Level trigger
Edge trigger: you need a change of status
o
Example
: Channel A1, rising edge, threshold = 0V: triggers only if the signal status
changes from negative to positive
Level trigger: no need to pass the threshold
o
Example:
Channel A1, high level, threshold = 0V: triggers only if the signal is
positive
Alarms only have level triggers.
Analogic / Logic channels
: trigger only from analogic or logic channels
Single / Multiple threshold
:
triggering from only one threshold,
complex triggering from several channels and several thresholds; see description below.
Such complex triggering is only possible with analogic channels.
You can setup these triggers thanks to the « trigger » menu in the homepage.