To maximize your chances of an accurate comparison between meter
and laboratory results, follow a few basic guidelines:
Before going to the lab
·
Perform a control solution test to make sure the meter is working properly.
·
Do Not eat for at least eight hours before you test your blood.
·
Take your meter with you to the lab.
While at the lab
Conduct your meter test within 15 minutes of the lab test.
·
Use fresh capillary blood obtained from the fingertip.
·
Follow all instructions in this Owners Booklet for performing a blood
·
glucose test with your meter.
Performance Characteristics
43
Precision:
Control Solution av 2.5mmol
/
L
Intermediate
Precision
Control Solution av 7.0mmol
/
L
Control Solution av 20.0mmol
/
L
Blood av 2.5mmol/L
Blood av 4.1mmol/L
Blood av 7.1mmol/L
Blood av 10.6mmol/L
Blood av 19.9mmol/L
Repeatability
Table3. System accuracy results for glucose between 2.2 mmol/L (40 mg/dL)
and 33.3 mmol/L (600 mg/dL)
This study shows that the system compares well with a laboratory method
and meets the minimum acceptable performance criteria defined in EN ISO
15197 : 2015.
600/600(100.0%)
Within ± 0.83 mmol/L or Within ±15%
SD
=
0.12mmol
/
L
CV
=
2.4%
CV
=
3.7%
SD
=
0.12mmol
/
L
CV
=
2.9%
CV
=
1.9%
CV
=
1.9%
CV
=
3.8%